B. Default Alarm Settings
The Dose and Rate alarm settings (thresholds)
for the UltraRadiac-Plus are factory set to the
values in the following tables.
R
Low Setting
High Setting
Rate
500 µR/hr
2 R/hr
Dose
100 mR
10 R
Sv
Low Setting
High Setting
Rate
1 mSv/hr
100 mSv/hr
Dose
1 mSv
100 mSv
The factory settings can be changed manually
(refer to the chapter
Setting Up the Unit
on
page 28), or via the optional Model S900
UltraRadiac Setup Software.
CAUTION:
The System Administrator is
responsible for determining
and setting the proper
operational alarm thresholds
for each UltraRadiac-Plus.
65