Chapter 4 Operational Verification Tests
Checking the Alarm
At any time during the Clr/Test sequence of
numbers:
•
Press the
RATE
key – the audio alarm
should sound and the Rate LED, the left
and one below the display, should turn
on.
•
Press the
DOSE
key – the audio alarm
should sound and the Dose LED, the
right hand one below the display,
should turn on.
•
Press the
ALARM
key – the audio
alarm should sound and both LEDs
should turn on.
•
If the optional vibrator alarm is
installed, the unit will vibrate when any
of these three keys is pressed.
Checking the Function Keys
When the
000
display appears during the
“Testing the Display” sequence, press each
function key in turn, except the
ON/OFF
key.
22