The Rate Mode
Reading the Rate Alarm Threshold
To display the High level (H) or Low level (L)
rate alarm thresholds:
•
Press the
ALARM
and
RATE
keys.
•
You’ll see an
H
, for High Level Alarm,
displayed.
•
To read the Low Level Alarm setting,
press the
RATE
key again; the display
will show an
L
.
•
Press the
CLR/TEST
key; the flashing
display will show the selected (
H
or
L
)
alarm threshold setting.
•
If the optional Vibrator Alarm has been
activated, the display will alternate
between the threshold value and a
vertically scrolling horizontal bar
(–
–
–).
Changing the Rate Alarm Indicators
To change the High level (H) or Low level (L)
alarm indicator settings:
•
Press the
ALARM
and
RATE
keys.
•
You’ll see an
H
, for High Level Alarm,
displayed.