Error Displays
Error Displays
Above 200 R/hr (2.0 Sv/h), and up to 600 R/h
(6.0 Sv/h), the upper three-segment bar (– – –)
on the display will begin flashing to indicate
that:
•
The dose rate exceeds the range of the
UltraRadiac-Plus.
•
The error might be greater than 20%.
•
The value of the displayed dose rate
(and consequently of the calculated
dose) is not accurate.
•
The “true rate” might be significantly
higher than the displayed value.
Above 600 R/h (6.0 Sv/h) the unit will not
increase dose-rate measurement and will keep
displaying 600 R/h (6.0 Sv/h).