Ending the Display Test
The unit will beep and display the indications
shown in the following table. Any other
numerical indication indicates a fault in that
key’s circuit.
Function
Indication
DOSE
002
ALARM
010
RATE
001
LIGHT
020
CLR/TEST
004
Ending the Display Test
At the end of the test sequence, the blinking
9
will appear, indicating that the Rate Mode is
operational. A blinking
0
indicates a failure;
contact your local Mirion Technologies
(Canberra) Service Center.
The unit will return to the Rate Mode in about
ten seconds.