![MingTai SUN12 Скачать руководство пользователя страница 9](http://html1.mh-extra.com/html/mingtai/sun12/sun12_operation-manual_1801431009.webp)
7
use UP and Down to change the data, then press START to quit and save the
setting.
Press SET to enter the advanced settings status, press SET again to set the custom
setting of the working cycle, like the Temperature, Sterilizing time..., use UP and
Down to change the data you want to set, then press START to quit and save the
setting.
9) Door Status light (auto-door lock only, manual door is invalid)
10) DOWN select key
When in program Setting, the down button were used to change the parameter
data down.
11)
PROGRAM key
Press PROGRAM, use UP and DOWN to select program, the temperature
window display sterilizing temperature of selected program; Pressure window
display sterilizing pressure of selected program; The Status window display
selected cycle code as below:
“
-1
”
: 134
℃
UNWRAPPED
“
-2
”
: 134
℃
WRAPPED
“
-3
”
: 121
℃
COTTON
“
-4
”
:121
℃
PLASTIC
“
-5
”
: 134
℃
PRION
“
-6
”
: Custom set cycle
“
-7
”
: BD test
“
-8
”
: Helix test
“
-9
”
: Vacuum test
Press START to confirm the selected program.
12)
START Key
Press the Starting button for starting the sterilizing process or save the setting.
7.2
Fill & drain water
When you install the Steam Sterilizer for the first time, after turn on the switch,
then The machine would displaying lack of the source water. Then you need to
use the measuring cup to fill distilled water from the top inlet around 1500ml for
the first time, next time do not filling the water too full.
Содержание SUN12
Страница 18: ...16 NO 2 Steam loop Work Description...