4)Note: The inner circle of the ring may be raised up when insert the ring into the
groove. Please press it to the groove by using the screwdriver carefully.
Shut off the power and cooling the Steam Sterilizer sufficiently before inserting
the ring for avoiding scald.
9.6 Check the safety valve
In order to prevent the safety valve in the blocking state, Under normal use, allow
the steam pressure release through it once a month.
1)operation sterilizing procedure
2)When the pressure in the sterilizer reaches 100 kpa
,
Pull rings on safety
valve, make the safety valve open,about 2S, if steam jet, the safety valv
e work is normal. If not, the safety valve need for effective inspection or
replacement
Pull rings on safety valve will steam jet, It is best to use tools such
as screwdriver when pull. Do not use finger directly, Operator also need ke
ep away from it , prevent burns!
Replace the safety valve:
This operation is limited to the professionals
1) Loosen the set screws of the safety valve, take the safety valve down f
rom the panel,
2) replace with a qualified safety valve (must provide by manufacturers).
3) Test a sterilization process
The machine maintenance and replacement of electrical components must cut
off power supply first. And only the manufacturer or professional person
designated by the manufacturer can operate
10.
Transportation and Storage
10.1 The preparatory work before transportation
and storage, Shut off the power, draw out the
plug and cool the autoclave down.
10.2 Draining water
Empty the water storing tank and the used water tank; The terminal side of the
attached pipe with no connecting joint shall be inserted into the water exiting joint
pipe. As the above picture showing, 3 is the water exiting mouth for used water
Содержание SUN12
Страница 18: ...16 NO 2 Steam loop Work Description...