13 

collecting tank, 4 is the water exiting mouth of the water storing tank. Revolving 
the water exit button by anticlockwise direction so as to release the water. 
 
10.3 The autoclave shall be transferred and stored with the below conditions: 
Ambiance temperature: -

40ºC 

-

+55ºC

 

Relative humidity: 

  85% 

Atmospheric pressure: 500HPA-1060HPA 

    Note: Don't drag or drop during transportation.   

 

11. Attention

 

 

ATTENTION 

1.

 

The Steam Sterilizer must be placed on the horizontal working base. 

2.

 

Must use the distilled water for the purpose of lasting the operation life of the 
Steam Sterilizer. 

3.

 

There shall be no clog to the air cooling window on the outer surface of the 
Steam Sterilizer. 

4.

 

The object apparatus to be sterilized shall be placed on the device plate with 
enough clearance retained in favor of the air circulation inside the  sterilizing 
room.   

5.

 

Empty the water inside the cooling water collecting tank frequently. Usually, 
the cooling water collecting tank shall be emptied once the water storing tank 
emptied. 

6.

 

The door handle must be buttoned up before working. 

7.

 

Don’t open the door

 of Steam Sterilizer before 

seeing the “0”

 kPa displayed 

on the pressure indicator. 

8.

 

Not  be  too  closed  to  the  door  of  the  Steam  Sterilizer  in  case  of  scald  when 
open the door. 

9.

 

Shut off the power before discharging/installing the sealing ring, and the work 
shall be carried out after further sufficient cooling in case of scald. 

10.

 

Don’t drag or drop the 

Steam Sterilizer during relocation period of the Steam 

Sterilizer. 

11.

 

The protecting ground shall be reliable. 

12.

 

Equipment must be used away from the magnetic field. 

13.The  device  end  of  life  must  be  for  processing  scrapped  in  accordance         
with the local environmental protection laws and regulations.

 

14.Such  as  power  failure  occurs  during  the  working,  need  to  disinfect  the 
equipment out, it must be open the door 2 minutes later. 
15.The equipment must be far away from flammable items. 
16. The equipment must be reliable grounding. 
17.

Must use the attached plug, do not change it. 

18.When  Choose  121 

  sterilization  procedure,  the  highest  temperature  in 

Содержание SUN12

Страница 1: ...ration Manual Please read the manual carefully operate it according to the manual strictly and avoid any hazard This manual applies to SUN8 SUN12 SUN16 SUN18 SUN23 M Steam Sterilizer File No MTQM C1 7...

Страница 2: ...3 5 Installation 3 6 Packing accessory 5 7 Display and keypad description 6 8 Emergency Condition 9 9 Maintenance and Service 11 10 Transportation and Storage 12 11 Attention 13 Appendix NO 1 The wor...

Страница 3: ...eliable product in sterilizing 1 Printer 2 USB 3 Used water outlet 4 Used water outlet 5 Power switch 6 Power source socket 7 Fuse 2 Application Scope This machine can be applied in sterilizing for th...

Страница 4: ...erilizer Diameter X Depth 8L 170mmX310mm Chamber size of the Steam Sterilizer Diameter X Depth 12L 200mmX360mm Chamber size of the Steam Sterilizer Diameter X Depth 16L 230mmX360mm Chamber size of the...

Страница 5: ...ACTURE AUTHORISED REPRESENTATIVE IN THE EUROPEAN COMMUNITY SERIAL NUMBER CONSULT INSTRUCTIONS FOR USE Earthing 5 Installation 5 1 Remove the packing Dismantle thin film open the packing get the equipm...

Страница 6: ...d the around articles shall be 10cm as retained and the top surface shall be 20cm retained Used to exhaust or drain 5 4 The 2 front feet of Steam Sterilizer can be adjusted the height please ensure th...

Страница 7: ...nz Zmax 0 114 OHM sind solche St rungen nicht zu erwarten Bitte kontaktieren Sie Ihre lokales Energieversorgungsunternehmen f r weitere Informationen PREPARATION BEFORE OPERATION Before operation plea...

Страница 8: ...laying the air pressure inside the chamber Unit KPa 4 Door lock indicator light It is the light for door closing well can press START to working or during the working cycle the door were not opened 5...

Страница 9: ...ture window display sterilizing temperature of selected program Pressure window display sterilizing pressure of selected program The Status window display selected cycle code as below 1 134 UNWRAPPED...

Страница 10: ...ocess open the vacuum selection valve exhaust valve normally on and start the vacuum pump 300 seconds later Close the vacuum selection valve exhaust valve normally on and the vacuum pump to stop pumpi...

Страница 11: ...3 1 3 Sterilizing time 5 5 20 20 25 Drying time 15 18 20 20 20 7 4 Emergency exit Long press the start key 5 s Machine will out of sterilization procedures meanwhile EC alarm 7 5 After sterilization u...

Страница 12: ...sing slowly Inspect the filter inside the chamber keep it clean and no jam 6 E6 long du Problem of door switch or door being opened when working Inspect door switch or if door handle were revolved to...

Страница 13: ...ooth cloth with distilled water to wipe the surface of the sealing ring or sealing cap gently The user shall discharge the sealing ring for further cleaning or replacement if the air leaking problem c...

Страница 14: ...screwdriver when pull Do not use finger directly Operator also need ke ep away from it prevent burns Replace the safety valve This operation is limited to the professionals 1 Loosen the set screws of...

Страница 15: ...ting tank shall be emptied once the water storing tank emptied 6 The door handle must be buttoned up before working 7 Don t open the door of Steam Sterilizer before seeing the 0 kPa displayed on the p...

Страница 16: ...nge of temperature and pressure may cause the liquid bottle burst so that endanger the safety of the person and equipment 22 The chloride ion is an important factor of stainless steel corrosion damage...

Страница 17: ...ing the working pressure pressure data In this status the chamber have pressure 4 Displaying the working time Sterilizing status In this status the working time and the count down figure are displayin...

Страница 18: ...16 NO 2 Steam loop Work Description...

Страница 19: ...ystem close the door then press START button to start Ningbo Mingtai Medical Instrument Co Ltd Yunzhen Road Industrial Zone Yunlong Town Yinzhou Ningbo City China Tel 86 0574 88320096 Fax 86 0574 8832...

Отзывы: