Must  wait  after  the  pressure  release  to 

“0

  Kpa

” 

,  then  open  the  door! 

  Even  the  pressure  is  0  Kpa,  there  is  still  hot  steam  inside  the  chamber, 

Operator need to take care it and prevent scald! 
 
 

Error code and resolving measures

 

NO  Code   

display 

Alarm Tone    Malfunction content 

Resolving Measures 

E1 

long "du" 
 

Fault of temperature   
sensor inside the chamber 

Inspect if anything effected the 
sensor or wire not connecting well 
or sensor was broken 

E2 

long "du" 

Pressure over 240 KPa 

Inspect pressure sensor or vacuum   
pump working 

E3 

long "du" 

Fault of temperature sensor outside 
the chamber 

Inspect the temperature sensor 
outside of the chamber 

E4 

long "du" 

Super high inner temperature when 
pressure rising 

Inspect inside temperature sensor 
or pressure sensor 

E5 

long "du" 

Pressure releasing slowly 

Inspect the filter inside the chamber 
keep it clean and no jam 

E6 

long "du" 

Problem of door switch or door 
being opened when working 

Inspect door switch or if door 
handle were revolved to the   
right MAX position 

E7 

long "du" 

Fault of temperature senor on steam 
generator 

Inspecting  the  temperature  sensor 
on steam generator 

E8 

long "du" 

Fail on pressure rising 

Inspect water flowing road and 
steam leaking. 


 

EH 

long "du" 

Steam generator not heating 

Inspect connecting wire or 
resistance of steam generator 

10 

EF 

long "du" 

Pressure not over than 0 KPa 

Inspect water pump/steam generator   

11 

EL 

long "du" 

Pressure not over than 20 KPa 

Inspect  if  electric  valve  causing 
steam leak, or replace water pump   

12 

Eo 

long "du" 

Pressure not over than 65 KPa 

Inspect if the water pump working 
weakly. Or replace it 

13 

E9 

long "du" 

Fail on keeping pressure and 
temperature 

Inspect the steam leaking 

14 

EC 

long "du" 

The process interrupted artificially 
 

Long press the start key, can unlock 
alarm. 
 

 
 

Содержание SUN12

Страница 1: ...ration Manual Please read the manual carefully operate it according to the manual strictly and avoid any hazard This manual applies to SUN8 SUN12 SUN16 SUN18 SUN23 M Steam Sterilizer File No MTQM C1 7...

Страница 2: ...3 5 Installation 3 6 Packing accessory 5 7 Display and keypad description 6 8 Emergency Condition 9 9 Maintenance and Service 11 10 Transportation and Storage 12 11 Attention 13 Appendix NO 1 The wor...

Страница 3: ...eliable product in sterilizing 1 Printer 2 USB 3 Used water outlet 4 Used water outlet 5 Power switch 6 Power source socket 7 Fuse 2 Application Scope This machine can be applied in sterilizing for th...

Страница 4: ...erilizer Diameter X Depth 8L 170mmX310mm Chamber size of the Steam Sterilizer Diameter X Depth 12L 200mmX360mm Chamber size of the Steam Sterilizer Diameter X Depth 16L 230mmX360mm Chamber size of the...

Страница 5: ...ACTURE AUTHORISED REPRESENTATIVE IN THE EUROPEAN COMMUNITY SERIAL NUMBER CONSULT INSTRUCTIONS FOR USE Earthing 5 Installation 5 1 Remove the packing Dismantle thin film open the packing get the equipm...

Страница 6: ...d the around articles shall be 10cm as retained and the top surface shall be 20cm retained Used to exhaust or drain 5 4 The 2 front feet of Steam Sterilizer can be adjusted the height please ensure th...

Страница 7: ...nz Zmax 0 114 OHM sind solche St rungen nicht zu erwarten Bitte kontaktieren Sie Ihre lokales Energieversorgungsunternehmen f r weitere Informationen PREPARATION BEFORE OPERATION Before operation plea...

Страница 8: ...laying the air pressure inside the chamber Unit KPa 4 Door lock indicator light It is the light for door closing well can press START to working or during the working cycle the door were not opened 5...

Страница 9: ...ture window display sterilizing temperature of selected program Pressure window display sterilizing pressure of selected program The Status window display selected cycle code as below 1 134 UNWRAPPED...

Страница 10: ...ocess open the vacuum selection valve exhaust valve normally on and start the vacuum pump 300 seconds later Close the vacuum selection valve exhaust valve normally on and the vacuum pump to stop pumpi...

Страница 11: ...3 1 3 Sterilizing time 5 5 20 20 25 Drying time 15 18 20 20 20 7 4 Emergency exit Long press the start key 5 s Machine will out of sterilization procedures meanwhile EC alarm 7 5 After sterilization u...

Страница 12: ...sing slowly Inspect the filter inside the chamber keep it clean and no jam 6 E6 long du Problem of door switch or door being opened when working Inspect door switch or if door handle were revolved to...

Страница 13: ...ooth cloth with distilled water to wipe the surface of the sealing ring or sealing cap gently The user shall discharge the sealing ring for further cleaning or replacement if the air leaking problem c...

Страница 14: ...screwdriver when pull Do not use finger directly Operator also need ke ep away from it prevent burns Replace the safety valve This operation is limited to the professionals 1 Loosen the set screws of...

Страница 15: ...ting tank shall be emptied once the water storing tank emptied 6 The door handle must be buttoned up before working 7 Don t open the door of Steam Sterilizer before seeing the 0 kPa displayed on the p...

Страница 16: ...nge of temperature and pressure may cause the liquid bottle burst so that endanger the safety of the person and equipment 22 The chloride ion is an important factor of stainless steel corrosion damage...

Страница 17: ...ing the working pressure pressure data In this status the chamber have pressure 4 Displaying the working time Sterilizing status In this status the working time and the count down figure are displayin...

Страница 18: ...16 NO 2 Steam loop Work Description...

Страница 19: ...ystem close the door then press START button to start Ningbo Mingtai Medical Instrument Co Ltd Yunzhen Road Industrial Zone Yunlong Town Yinzhou Ningbo City China Tel 86 0574 88320096 Fax 86 0574 8832...

Отзывы: