9. Maintenance and Service 

 

Shut  off  the  power  before  service  or  components’  replacement,  and  the 

service or replacement work shall be done by the technician. 
 
Working environment minimum brightness to 215 + 15 lx. 
 
9.1 Clean the top water tank by disinfectant or alcohol once a week. 
9.2  Disinfect  and  clean  the  interior  surface  of  the  Steam  Sterilizer  chamber 
monthly. 
9.3 Replace the fuse.   

Fuse Type: 250V,15A 

  Check the parameters of new fuse is correct . 

1)

 

Before change the fuse, turn off the power, and must pull out the plug. 

2)  Revolve  the  fuse  base  by  counter-clockwise  by  screwdriver  to  bring  out  the 
melting fuse to be replaced; 
3) Replace the new fuse and reset the fuse base on the original position, then use 
the screwdriver to fasten the base by clockwise. 
4)Check the correctness of the data of the new fuse replaced.   
 
9.4 Clean the sealing ring periodically 
The operator shall clean the sealing ring periodically for avoiding the influence to 
the  seal caused  by  the  dust/dirt  left  due to  the  long  period  use.  Use  the  smooth 
cloth  with  distilled  water  to  wipe  the  surface  of  the  sealing  ring  or  sealing  cap 
gently.  The  user  shall  discharge  the  sealing  ring  for  further  cleaning  or 
replacement if the air-leaking problem can not be resolved after the above process 
(The user shall discharge the sealing ring periodically for cleaning and inspection 
as recommendation by the supplier).       
 
9.5 The replacement work for sealing ring 
Tool: One flat screwdriver (No sharp edge at the head) 
1)  One  hand  catches  the  lip  of  the  sealing  ring,  the  other  hand  insert  the 
screwdriver into the clearance between the sealing ring and the door for  raise up 
the sealing ring. 
2)After raising up part of the ring, you can use hand to draw the whole ring out. 

Wash its groove after drawing out the ring and pay attention to see whether it’s 

spoiled or not in consideration of the necessity of replacement. 
3)Put back the ring to the original door groove after cleaning. Most Important: the 
inset  work  must  be  done  equably to  the groove  During  the  installation,  the  laid 
four  equal  points  of  the  ring  must  be inserted  to the  groove  firstly, later  for  the 
same  work  for  the  remaining  segments  of  the  ring.  After  finishing,  press  the 
sealing ring equally by the force of hand. 

Содержание SUN12

Страница 1: ...ration Manual Please read the manual carefully operate it according to the manual strictly and avoid any hazard This manual applies to SUN8 SUN12 SUN16 SUN18 SUN23 M Steam Sterilizer File No MTQM C1 7...

Страница 2: ...3 5 Installation 3 6 Packing accessory 5 7 Display and keypad description 6 8 Emergency Condition 9 9 Maintenance and Service 11 10 Transportation and Storage 12 11 Attention 13 Appendix NO 1 The wor...

Страница 3: ...eliable product in sterilizing 1 Printer 2 USB 3 Used water outlet 4 Used water outlet 5 Power switch 6 Power source socket 7 Fuse 2 Application Scope This machine can be applied in sterilizing for th...

Страница 4: ...erilizer Diameter X Depth 8L 170mmX310mm Chamber size of the Steam Sterilizer Diameter X Depth 12L 200mmX360mm Chamber size of the Steam Sterilizer Diameter X Depth 16L 230mmX360mm Chamber size of the...

Страница 5: ...ACTURE AUTHORISED REPRESENTATIVE IN THE EUROPEAN COMMUNITY SERIAL NUMBER CONSULT INSTRUCTIONS FOR USE Earthing 5 Installation 5 1 Remove the packing Dismantle thin film open the packing get the equipm...

Страница 6: ...d the around articles shall be 10cm as retained and the top surface shall be 20cm retained Used to exhaust or drain 5 4 The 2 front feet of Steam Sterilizer can be adjusted the height please ensure th...

Страница 7: ...nz Zmax 0 114 OHM sind solche St rungen nicht zu erwarten Bitte kontaktieren Sie Ihre lokales Energieversorgungsunternehmen f r weitere Informationen PREPARATION BEFORE OPERATION Before operation plea...

Страница 8: ...laying the air pressure inside the chamber Unit KPa 4 Door lock indicator light It is the light for door closing well can press START to working or during the working cycle the door were not opened 5...

Страница 9: ...ture window display sterilizing temperature of selected program Pressure window display sterilizing pressure of selected program The Status window display selected cycle code as below 1 134 UNWRAPPED...

Страница 10: ...ocess open the vacuum selection valve exhaust valve normally on and start the vacuum pump 300 seconds later Close the vacuum selection valve exhaust valve normally on and the vacuum pump to stop pumpi...

Страница 11: ...3 1 3 Sterilizing time 5 5 20 20 25 Drying time 15 18 20 20 20 7 4 Emergency exit Long press the start key 5 s Machine will out of sterilization procedures meanwhile EC alarm 7 5 After sterilization u...

Страница 12: ...sing slowly Inspect the filter inside the chamber keep it clean and no jam 6 E6 long du Problem of door switch or door being opened when working Inspect door switch or if door handle were revolved to...

Страница 13: ...ooth cloth with distilled water to wipe the surface of the sealing ring or sealing cap gently The user shall discharge the sealing ring for further cleaning or replacement if the air leaking problem c...

Страница 14: ...screwdriver when pull Do not use finger directly Operator also need ke ep away from it prevent burns Replace the safety valve This operation is limited to the professionals 1 Loosen the set screws of...

Страница 15: ...ting tank shall be emptied once the water storing tank emptied 6 The door handle must be buttoned up before working 7 Don t open the door of Steam Sterilizer before seeing the 0 kPa displayed on the p...

Страница 16: ...nge of temperature and pressure may cause the liquid bottle burst so that endanger the safety of the person and equipment 22 The chloride ion is an important factor of stainless steel corrosion damage...

Страница 17: ...ing the working pressure pressure data In this status the chamber have pressure 4 Displaying the working time Sterilizing status In this status the working time and the count down figure are displayin...

Страница 18: ...16 NO 2 Steam loop Work Description...

Страница 19: ...ystem close the door then press START button to start Ningbo Mingtai Medical Instrument Co Ltd Yunzhen Road Industrial Zone Yunlong Town Yinzhou Ningbo City China Tel 86 0574 88320096 Fax 86 0574 8832...

Отзывы: