7.3.4 Exhaust drying process:   

When the sterilization time, reduced to zero, into the exhaust stroke dry, open the 
drain valve to drain the pressure reduced to +0.0then the vacuum pump to start. 
When the drying time only 3 minutes ,open the intake valve to vacuum drying.   

7.3.5 Pressure equilibrium process:   

When the drying time is reduced to zero, turn off the exhaust, dry intake valve 
opens, so that the pressure to achieve and external balance, open the automatic 
doors, complete sterilization.   

7.3.6 Sterilization program is as follows:

 

 

 

 

134

 

UNWRAPPED 

134

 

WRAPPED 

121

 

COTTON 

121

 

PLASTIC 

134

 

PRION 

Sterilizing 
temperature 

134

 

134

 

121

 

121

 

134

 

Sterilizing 
pressure 

210KPa 

210KPa 

110KPa 

110KPa 

210KPa 

Vacuum times 

Sterilizing  
time 

20 

20 

25 

Drying time 

15 

18 

20 

20 

20 

 
7.4 Emergency exit, Long press the start key 5 s, Machine will out of sterilization procedures, 

meanwhile 

"EC" alarm 

7.5 After sterilization, use manual holder to take the sterilized devices out. 

  Must use manual holder to take the device,and prevent scald! 

 

8. Emergency Condition 

During  the  working  time,  the  Steam  Sterilizer  will automatically  alarming, then 
release  pressure,  stop  heating  and  display  the  error  code  (The  definition  of  the 
error  code  is  as  the  below  table:  error  code  and  resolving  method  ).  It  would 
ensure the safety of the operator if the emergency occur. 
Any  emergency  happen,  please  check  the  warning  code  table  for  resolving  the 
according malfunctions. If not possible to resolve, please contact the dealers. We 
will try our best to provide help to you in short time. 
 
  After  the  alarm, The machine  can't  work,  Long  press  the  start  key,  can  unlock 
alarm. 

Содержание SUN12

Страница 1: ...ration Manual Please read the manual carefully operate it according to the manual strictly and avoid any hazard This manual applies to SUN8 SUN12 SUN16 SUN18 SUN23 M Steam Sterilizer File No MTQM C1 7...

Страница 2: ...3 5 Installation 3 6 Packing accessory 5 7 Display and keypad description 6 8 Emergency Condition 9 9 Maintenance and Service 11 10 Transportation and Storage 12 11 Attention 13 Appendix NO 1 The wor...

Страница 3: ...eliable product in sterilizing 1 Printer 2 USB 3 Used water outlet 4 Used water outlet 5 Power switch 6 Power source socket 7 Fuse 2 Application Scope This machine can be applied in sterilizing for th...

Страница 4: ...erilizer Diameter X Depth 8L 170mmX310mm Chamber size of the Steam Sterilizer Diameter X Depth 12L 200mmX360mm Chamber size of the Steam Sterilizer Diameter X Depth 16L 230mmX360mm Chamber size of the...

Страница 5: ...ACTURE AUTHORISED REPRESENTATIVE IN THE EUROPEAN COMMUNITY SERIAL NUMBER CONSULT INSTRUCTIONS FOR USE Earthing 5 Installation 5 1 Remove the packing Dismantle thin film open the packing get the equipm...

Страница 6: ...d the around articles shall be 10cm as retained and the top surface shall be 20cm retained Used to exhaust or drain 5 4 The 2 front feet of Steam Sterilizer can be adjusted the height please ensure th...

Страница 7: ...nz Zmax 0 114 OHM sind solche St rungen nicht zu erwarten Bitte kontaktieren Sie Ihre lokales Energieversorgungsunternehmen f r weitere Informationen PREPARATION BEFORE OPERATION Before operation plea...

Страница 8: ...laying the air pressure inside the chamber Unit KPa 4 Door lock indicator light It is the light for door closing well can press START to working or during the working cycle the door were not opened 5...

Страница 9: ...ture window display sterilizing temperature of selected program Pressure window display sterilizing pressure of selected program The Status window display selected cycle code as below 1 134 UNWRAPPED...

Страница 10: ...ocess open the vacuum selection valve exhaust valve normally on and start the vacuum pump 300 seconds later Close the vacuum selection valve exhaust valve normally on and the vacuum pump to stop pumpi...

Страница 11: ...3 1 3 Sterilizing time 5 5 20 20 25 Drying time 15 18 20 20 20 7 4 Emergency exit Long press the start key 5 s Machine will out of sterilization procedures meanwhile EC alarm 7 5 After sterilization u...

Страница 12: ...sing slowly Inspect the filter inside the chamber keep it clean and no jam 6 E6 long du Problem of door switch or door being opened when working Inspect door switch or if door handle were revolved to...

Страница 13: ...ooth cloth with distilled water to wipe the surface of the sealing ring or sealing cap gently The user shall discharge the sealing ring for further cleaning or replacement if the air leaking problem c...

Страница 14: ...screwdriver when pull Do not use finger directly Operator also need ke ep away from it prevent burns Replace the safety valve This operation is limited to the professionals 1 Loosen the set screws of...

Страница 15: ...ting tank shall be emptied once the water storing tank emptied 6 The door handle must be buttoned up before working 7 Don t open the door of Steam Sterilizer before seeing the 0 kPa displayed on the p...

Страница 16: ...nge of temperature and pressure may cause the liquid bottle burst so that endanger the safety of the person and equipment 22 The chloride ion is an important factor of stainless steel corrosion damage...

Страница 17: ...ing the working pressure pressure data In this status the chamber have pressure 4 Displaying the working time Sterilizing status In this status the working time and the count down figure are displayin...

Страница 18: ...16 NO 2 Steam loop Work Description...

Страница 19: ...ystem close the door then press START button to start Ningbo Mingtai Medical Instrument Co Ltd Yunzhen Road Industrial Zone Yunlong Town Yinzhou Ningbo City China Tel 86 0574 88320096 Fax 86 0574 8832...

Отзывы: