9
4−3. Connections
4−3−1. Connection with strain gage applied transducers
The instrument can connect with strain gage applied transducers, such as load cell, pressure
transducer and so on. Here, we will describe the example of connections with load cell, so
connect with another type of strain gage applied transducers in the same way.
※
1 When tension is applied with the application of tension type or
universal(compression/tension) type of load cell, and display of “+”
direction is required, connect “Green” with Terminal No.9 at
terminals 1, and “blue” with Terminal No.8 at terminals 1
individually. As there is a case which standard wiring color is
different, please confirm the inspection data sheet of the load cell
being used.
※
2 When the total length of CAB−502 specifies more than 30 m, the
accuracy may be out of warranty because the resistance of cable
makes the input voltage of the instrument decreased.
※
3 When the total length of CAB−501 specifies more than 100 m, the
accuracy may be out of warranty, because the resistance of cable
makes the remote sensing function of the instrument deteriorated.
1
Connection with one set of Load cell and CSD−581−15
3 m
RED
Enlarged figure for connecting section
RED
WH
T
※
1G
R
N
※
1B
L
U
Sh
ie
ld
Attached short bars
WH
T
※
1B
L
U
Sh
ield
Attached short bars
※
1G
R
N
CN202