21
5
−
2
−
1. Calibration method to register the output of strain gage applied transducer at the time
of maximum display after setting the load to zero.
Warning
●
Before using a new instrument or exchanging the strain gage applied
transducer for a new one, be sure to make calibration.
If calibration shall not be made, correct measured results may not be
obtained nor may cause malfunction in the instrument and there
may exist damage in peripheral equipments.
Besides, even though calibration had been made, there may occur the
similar case when the result is not correct, so make calibration again.
●
During the calibration is executing, be sure to set Tare weight
cancellation clear, and to make cancellation (Execution of F
−
98) for
compensated data on zero set and also set the OFF position of Zero
tracking (Setting “00000” on F
−
07 and F
−
08 each.) If neglected,
correct measured results may not be obtained.
●
During calibration procedures, press the “FUNC.” key in case of
interrupting the calibration is required. The calibration data will be kept as
they are before entering the calibration and then returns to the
Measurement mode.
●
When the “A/Z
A
” key is pressed with the load display of “FUNC.”,
the display will change as the following sequence at every time the
key is pressed. “FUNC”
→
“CCAL”
→
“ACAL”
→
“LCAL”
→
“ZERO”
→
“SPAN”
→
“TARE”
→
“CHECK”
→
“MONIT”
→
“F
−
END”
→
“FUNC”
→
“CCAL”
→・・・・・
(Hereinafter, it will repeat.)
Procedures
1
Press the “FUNC.” key for approx. one second.
The load display will show “FUNC”.