Set the connection range to DC
・
V for the measuring instrument such as tester and
so on.
①
Remove the connecting cable from the terminals at the strain gage applied transducer.
②
Measure the voltage between terminal A and terminal C on the terminals.
Set the connecting range to DC
・
mV on the measuring
instrument such as tester and so on.
①
Re
−
connect the connecting cable with the terminals on the strain
gage applied transducer. (Refer to the chapter 4.)
②
Checking the voltage between the terminal B and D on the terminal
board, apply load on strain gaga applied transducer.
The voltage between B
and D varies.
The voltage between
A and C is just like
the setting of SW1.
6
Inform
us
about
the
contents
of
failure
and
situation
at
site
in
details.
Contact
us
.
Make calibration again according to the
procedures in the chapter 5.
Start of measurement
Same condition
There may be the fear that strain gage
applied transducer connected has
destroyed or signal route (wire) has cut off.
YES
NO
YES
YES
NO
NO