65
●
When Tare weight cancellation (A/Z) is executed in effective condition
of digital Tare weight cancellation, the display value will be shown as
follows:
(Indication when Tare weight cancellation in effective condition of
digital tare weight cancelation)
=(
Gross weight
)−(
Tare weight cancellation value)
7
−
11. Key lock
The instrument prepares key lock function.
With the Function F
−
06, each digit setting =0 execute key lock OFF , and each digit of setting =1
executes key lock ON. As for default, all are key lock OFF.
Besides, the correspondence between the target of key lock and setting digits are as follows:
10
0
digit
:
CHECK
10
1
digit
:
FUNC.
10
2
digit
:
ZERO
10
3
digit
:
A/Z OFF
Y
10
4
digit
:
A/Z
A
Moreover, when the “FUNC.” key is key locked , lock on “FUNC.” key will be cancelled only
once after the “FUNC.” key is pressed together with the “ENTER” key for more than 3 s.
7
−
12. CHECK value
When the “CHECK” key is pressed in the Measurement mode, CHECK value that are equivalent to
the set value with the Function F
−
43 will be ON. And at the same time when the “CHECK” on
status display lights up, the load value equivalent to the set value with Function F
−
43 will be
added to the load display value.
By pressing the “CHECK” key again, the CHECK value will be OFF and returns to the former
condition as it is. Moreover, when the CHECK value is ON, “RUN” output will be OFF and status
display of “RUN” will be put off, however, output of BCD will keep on outputting as same in the
Measurement mode.
●
When the setting of Function F
−
43 is “00000”, load display value will
not vary even the “CHECK” lights up on the status display by
pressing the “CHECK” key.
(
Because the CHECK is 0.0mV/V.)