41
5
−
2
−
6. Calibration procedure to apply registration again for zero point only
Warning
●
At the time of executing the calibration, be sure to set tare weight
cancellation clear, cancellation of compensated data at zero set
(Execution of F
−
98) and zero tracking OFF(Setting of F
−
07, and
F
−
08 to “00000”). If neglected, there may be the case that correct
calibration is not executed.
●
During the calibration procedure, press ”FUNC.” key to interrupt the
calibration. It will hold the calibration data before interrupting and
will return to the Measurement mode.
●
When the “A/Z
A
” key is pressed with the load display of “FUNC.”,
the display will change as the following sequence at every time the
key is pressed. “FUNC”
→
“CCAL”
→
“ACAL”
→
“LCAL”
→
“ZERO”
→
“SPAN”
→
“TARE”
→
“CHECK”
→
“MONIT”
→
“F
−
END”
→
“FUNC”
→
“CCAL”
→・・・・・
(Hereinafter, it will repeat.)
Procedures
1
Press the “FUNC.” for about one second.
The load display will show “FUNC”.
2
Press the “A/Z
A
” key 6 times.
It will make the load display as “CCAL”
→
“ACAL”
→
“LCAL”
→
“ZERO”
→
“SPAN”
→
“TARE”.
Here, set to apply the initial load on it.