— CAL
— LIN
Selection of the scale to be linearized and
starting the linearization initiates a linearization
process for WP1
— WP 1
Initiates a linearization process for WP1
This process can be canceled by pressing the
(Exit) key.
The weight values are fixed.
— WP 2
Only visible if a second scale is defined in the
menu
Initiates a linearization process for WP2.
This process can be canceled by pressing the
(Exit) key.
The weight values are fixed.
— GEO
— END
Exit menu level
5.4.2.2
Menu settings
CAL.ADJ
5.4.3
Geographic adjustment factor (GEO) – procedure
The adjustment of the calibration based on the current location is carried out using the
geographic adjustment factor [GEO]. (Settings from 0 … 31 are available.) The table under
contains the GEO values for a wide range of latitudes.
Call up the menu mode:
Weight
Count
First weight value (flashing)
1.
Press and hold the
key.
Weight
The first menu item [METRO] is displayed in the "Weight" display.
2.
Press the
(No) key until the menu item [CAL.ADJ] is displayed in the "Weight"
display.
3.
Press the
(Yes) key to access the sub-menu item.
Compact scale Puro® LargeCount
5 Operating
EN-58
Minebea Intec