![Minebea Intec Puro LargeCount EF-LT P1 Скачать руководство пользователя страница 58](http://html1.mh-extra.com/html/minebea-intec/puro-largecount-ef-lt-p1/puro-largecount-ef-lt-p1_operating-instructions-manual_1801213058.webp)
Adopt the second weighing point. During the internal calculations, the same display
appears in the display as for step 2.
If the calculations were error-free, the following is displayed on the display for
around 2 seconds:
Weight
Second menu level = calibration
Piece Weight
Third menu level = here WP1
Count
OK message = calibration is complete
The device exits the menu and returns back to the normal application mode.
If an error occurred during the calibration, the following is displayed on the displays
for around 2 seconds:
Weight
Second menu level = calibration
Piece Weight
Third menu level = here WP1
Count
Error message = calibration is canceled
Piece Weight
Third menu level = here WP1
Count
Flashing weight, which can be changed
via the keypad
3.
If the correct second weighing point is selected, press the
(Yes) key.
Compact scale Puro® LargeCount
5 Operating
EN-56
Minebea Intec