If the unit used for normal weighing is ounces, the calibration unit is also oz (ounce).
If a metric unit (kg or g) is used for normal weighing, the calibration unit is kg (kilograms).
5.4.1.1
Calibration
1.
WP1 or WP2 (if available) has been activated.
Weight
Second menu level = calibration
Piece Weight
Third menu level = here WP1
Count
First weight value (flashing)
The following is displayed in the displays:
2.
Remove all weights from the weighing pan and press the
(Yes) key.
In this time, the following is displayed in the displays:
Weight
Second menu level = calibration
Piece Weight
Third menu level = here WP1
Count
Wait message, until the weight has been
adopted
The second weighing point is displayed flashing.
Weight
Second menu level = calibration
Now the first calibration point is adopted and saved internally.
5 Operating
Compact scale Puro® LargeCount
Minebea Intec