The basic indicator is factory calibrated for both the high and low mV input

ranges. In the application, the user scales the basic mV to units of weight (Input
Scaling). To re-calibrate the complete weighing system (load cells and indicator
together) with calibration weights, execution of the Input Scaling procedure
suffices. It generally is not necessary to re-calibrate the input of basic indicator.

In the case of using known data from load cells where the mV readings are

directly keyed-in, then the input can be re-calibrated. Calibration reference
accuracy of 0.01% in the 20 and 200 mV ranges is required.

Calibration of the indicator is divided into two parts: the Input Calibration

and Analog Output calibration.

Input Calibration

Note: This procedure affects the basic indicator mV input calibration. As a

result, it may require the indicator to be rescaled using calibration weights.

From the 

40/

display, select the input or output for calibration:

1"15

20 mV input (Set 20 mV range jumper)

1"5

200 mV input (Set 200 mV range jumper)

Output Calibration

The Analog Output signal is re-calibrated by setting the end points of each

range. These end points correlate with Analog Output scaling parameters AN-
LO and AN-HI. Calibration meter accuracy of 0.05% in the 10 VDC and 20 mA
ranges is required. 

9

PARAMETER MODULE 8 - Analog Output Parameters

PARAMETER MODULE 7 - Serial Communications Parameters

PARAMETER MODULE 9 - Factory Service Function Parameters

DISPLAY

PARAMETER

RANGE AND UNITS

DESCRIPTION/

COMMENTS

2400

baUd

Baud Rate

600

Set baud rate to match
other equipment.

Set to 9600 for PC
configuration software.

9600

19200

abrv

addr

par

data

Abbreviated
Output

Node Address

Parity Bit

Number of
Data Bits

EVEN

8

8 = No parity

300

1200

4800

NO

NO

- ID characters printed

0

to 

99

Odd

7

DISPLAY

PARAMETER

RANGE AND UNITS

DESCRIPTION/

COMMENTS

0-10

- 0 to 10 V

typE

Select Analog
Output Type

4-20

- 4 to 20 mA

asIN

Output Source
Assignment

*

HI

- Max reading

Select source of linear
output signal.
If totalizer, scaling
limited to lower 5 digits.

*

tot

- Totalizer value

0-20

- 0 to 20 mA

NEt

- Net weight

b-6

- Brutto/Gross weight

*

LO

- Min reading

YES

- Enter sub-menu

t

- Tons

NO

- None

NO

- Disables parameter

0

- False zero

6

- Grams

Print Options

Line Print Units
Suffix

Opt

UNIt

*

tot

- Totalizer values

0oc

- Zero and ounces

**

6roSS

- Brutto/Gross weight

06

- Zero and grams

K6

- Kilograms

SPNt

- Setpoint values

NEt

- Net weight

0Lb

- Zero and pounds

oc

- Ounces

**

tArE

- Tare weight

0K6

- Zero and kilograms

lb

- Pounds

DESCRIPTION/

COMMENTS

RANGE AND UNITS

PARAMETER

DISPLAY

Sub-menu to select
parameters for inclusion
on print ticket.

Select weight units for
line print.

Select full or abbreviated
transmission

Possible Combinations:
8, N, 1
7, O, 1
7, E, 1
7, N, 2

Display Scale
Value High

AN-HI

Low = 0 mA/ 4 mA/ 0 V
High = 20 mA/ 10 V

-19999

to 

99999

Display Scale
Value Low

AN-LO

0.0 = Instant updates

0.0

to 

10.0

sec

Update Time

DESCRIPTION/

COMMENTS

RANGE AND UNITS

PARAMETER

DISPLAY

For reverse acting,
reverse scale high and
scale low

udt

*

HILO

- Max and Min values

Unique address for each
indicator. Set to 0 for PC
configuration software.

YES

- ID characters suppressed

* - Not in “B-GNT” mode

** - Not in “HILOT” mode.

* - Not in “B-GNT” mode

DISPLAY

ACTION

1"116

or

1"16

Apply 0 mV b and - SIG inputs, with -SIG tied
to common, wait 10 seconds, press PAR

1"156

or

1"56

Apply +20 or +200 mV to + and - SIG inputs, with -SIG
tied to common, wait 10 seconds, press PAR

DISPLAY

ACTION

0.0_A

4.0_A

20.0_A

0.0v

10.0v

Use Tare and Zero keys to set mA output to 0.00 mA. 
Press PAR to accept.

Use Tare and Zero keys to set mA output to 20.00 mA.
Press PAR to accept.

Use Tare and Zero keys to set mA output to 4.00 mA.
Press PAR to accept.

Use Tare and Zero keys to set volt output to 0.00 V. 
Press PAR to accept.
Use Tare and Zero keys to set volt output to 10.00 V. 
Press PAR to accept.

0 = dimmest
15 = brightest

to

15

Display Intensity
Level

d-LEv

All other codes result in
exit.

48

- Calibration

66 

- Restore Factory

Defaults

Factory Service
Access Code

CodE

DESCRIPTION/

COMMENTS

RANGE AND UNITS

PARAMETER

DISPLAY

Содержание MP30

Страница 1: ...uts implemented on Plug in cards The Plug in cards provide dual FORM C relays 5 A quad FORM A relays 3 A or either sinking or sourcing quad open collector logic outputs The setpoint alarms can be configured in modes to suit a variety of control and alarm requirements High and low absolute high and low deviation and band acting Balanced or unbalanced hysteresis On and off delay timers Auto reset or...

Страница 2: ...RS485 Isolation To Sensor User Input Commons 500 Vrms for 1 min Working Voltage 50 V Not Isolated from all other commons Data 7 8 bits Baud 300 to 19200 Parity no odd or even Bus Address selectable 0 to 99 Max 32 meters per line RS485 Transmit Delay Selectable for 2 to 50 msec or 50 to 100 msec RS485 17 ANALOG OUTPUT Types 0 to 20 mA 4 to 20 mA or 0 to 10 VDC Isolation To Sensor User Input Commons...

Страница 3: ... parameter and indexes to the next During Digital Tare this key stores the tare weight value keyed in KEY The Tare In Tare Out key both acquires and clears the Tare weight value The function of this key can be changed by setting the F1 function parameter For data entry this key is also used to increment a parameter value KEY The key Zero re zeros the Basic and Gross readout displays The function o...

Страница 4: ...as a basic millivoltmeter The basic indicator readout can then be post scaled to read out in weight units The indicator provides two ways in which to scale the display Key in Key in the input and display scaling points using known data Apply Apply the actual input value and key in the corresponding display value The indicator records the input value applied For processes that require linearity com...

Страница 5: ...ons Each time the batch function is executed the weight is added to the total and then zeroed PARAMETER LOCK MODE A user input can be used to lock the parameter list When the user input is active the indicator is in the protected parameter mode where it is only possible to access the setpoint values and the security code It is possible to lock the parameter list without using a user input as a pro...

Страница 6: ...deviation or band acting the effective trigger point is offset by SP1 In these modes SP1 acts as a master setpoint that the other setpoints track An alternate list of setpoint values can be stored and recalled as needed Select the alternate list with a function key or user input and program the setpoint values This allows for quick changeover for a new process condition Setpoint Alarm Hysteresis a...

Страница 7: ...eset Active Reset LIst Select main or alternate setpoint list r 1 Reset setpoint 1 r 2 Reset setpoint 2 USr 2 USr 1 User Input Function no no function pLOC Lock parameter list USr 3 rEL Acquire zero Settings 0 3 8 15 Active Reset Active Function Active Min Active Function Active Max Active Enable Active Enable Active Reset Active Total Active edge Batch Release from active Synchronize Active Hold ...

Страница 8: ...ale Factor sCfaC DESCRIPTION COMMENTS RANGE AND UNITS PARAMETER DISPLAY Tracking on reading band Tracking off reading 0 reading band PARAMETER MODULE 6 Setpoint Parameters DISPLAY PARAMETER RANGE AND UNITS DESCRIPTION COMMENTS act n Action for Setpoint ab HI absolute high balanced hysteresis SPSEL Select Setpoint Select setpoint for configuration Select operating action for setpoint See Figures fo...

Страница 9: ...sIN Output Source Assignment HI Max reading Select source of linear output signal If totalizer scaling limited to lower 5 digits tot Totalizer value 0 20 0 to 20 mA NEt Net weight b 6 Brutto Gross weight LO Min reading YES Enter sub menu t Tons NO None NO Disables parameter 0 False zero 6 Grams Print Options Line Print Units Suffix Opt UNIt tot Totalizer values 0oc Zero and ounces 6roSS Brutto Gro...

Страница 10: ...ers 5 tOt Totalizer Parameters 8 Out Analog Output Parameters 7 SrL Serial Communication Parameters DISPLAY PARAMETER USER SETTING FACTORY SETTING B 6 tArE BRUTTO GROSS DISPLAY TARE DISPLAY rEd Ent SP 3 SP 1 SETPOINT 3 SETPOINT 1 LOC LOC SP 4 SP 2 SETPOINT 4 SETPOINT 2 LOC LOC CodE SECURITY CODE 00000 OPEr HI t MOPERATING MODE MAX CAPTURE DELAY TIME b 6Nt 0000 0 LO t MIN CAPTURE DELAY TIME 0000 0 ...

Страница 11: ...rivers pens pencils etc to operate the keypad of the unit EMC GUIDELINES The indicator has been designed and tested to meet the requirements of EMC Directive 89 336 EEC However successful installations depend upon several factors These include cable routing termination of cable shields associated equipment etc Installation The indicator meets NEMA 4X IP65 requirements for indoor use when properly ...

Страница 12: ...med as a low acting alarm triggering on the Net liquid weight Each filling operation is started with a Tare Out Tare In sequence This opens the filling valve and when the target setpoint is reached the valve is shut off The valve remains off until a tare operation is repeated thereby restarting the filling operation The fourth output is connected in series with the fill output to disable filling d...

Отзывы: