MANUAL ZERO ACQUISITION

In Manual Zero operation, the Gross weight is zeroed each time the function

is executed. The function is activated by a key press or user input transition. If
the total accumulated zero offset exceeds the Zero Acquisition Range, the
function is disabled. In this case, the indicator must be re-calibrated to the load
sensing device, in order to reset the true system zero and hence the accumulated
zero offset. The accumulated zero offset itself is reset whenever any of the input
scaling parameters are altered.

The indicator displays a brief message 

rEsEt

whenever a zero operation

occurs. The indicator displays 

NO

whenever the total accumulated zero exceeds

the value of the Zero Acquisition parameter. In this case, the zero operation is
blocked.

“B-GNT” Operating Mode

In the “

b-6Nt

” operating mode, the Basic, Gross, Net and Tare Weight can be

displayed and controlled. In this mode, indicator functions that pertain to Max
and Min reading memory and the Totalizer are disabled. This is the default mode.

“HILOT” Operating Mode

In the “

HILOt

” operating mode, the Max, Min and Totalizer indicator

functions can be displayed and controlled. In this mode, the Gross, Net and Tare
weight displays and control are disabled (Basic display only.). The table lists the
correspondence of the display annunciators in this mode.

*

Only functional in “HILOT” operating mode.

SETPOINT ALARMS

The indicator has two setpoint alarm Plug-in cards. Only one of these cards

can be installed at a time. These are: 

Dual relay, FORM-C
Quad relay, FORM-A

The setpoint alarms can be independently configured for a variety of operating

modes. It is also possible to set the alarms to trigger from the totalizer value.
Each setpoint alarm can be programmed to trigger from the Brutto/Gross weight
or the Net weight. The setpoint alarm figures illustrate the operating modes:

Setpoint Alarm Values

When setpoint alarm is programmed as deviation or band acting, the effective

trigger point is offset by SP1. In these modes, SP1 acts as a master setpoint, that
the other setpoints track.

An alternate list of setpoint values can be stored and recalled as needed.

Select the alternate list with a function key or user input and program the
setpoint values. This allows for quick changeover for a new process condition.

Setpoint Alarm Hysteresis and On/Off Time Delay

Depending on the application, hysteresis or On/Off time delays can be used.

When the setpoint is a control output, hysteresis is usually used to control the
cycle period. Optionally, time delay can also be employed to ensure minimum
On/Off times. This is useful to limit the cycling of mechanical control equipment.

On/Off time delays are effective when the setpoint is used as an alarm output.

The time delays can be used to prevent false triggering during process transient
events, while hysteresis eliminates output chatter at the switch point.

The hysteresis value can be selected to act in the balanced mode or

unbalanced mode. In the balanced mode, the hysteresis action is centered about
the trigger point, while in the unbalanced mode the hysteresis acts on one side
of the trigger point. When the setpoint is a control output, usually balanced
hysteresis is used. This tends to center the oscillation around the setpoint value.
For alarm applications, usually unbalanced hysteresis is used. This makes the
alarm activate precisely at the trigger value.

Setpoint Alarm Output Phase and Panel Annunciator

The output phase (logic) of each setpoint can be reversed. While the front

panel indicator follows the output state, it can be programmed to reverse as
well; or flash for added emphasis.

Setpoint Alarm Reset Modes, Standby Operation 

Each setpoint alarm can be programmed for auto or latching reset modes.

Standby operation can be used to prevent the triggering of a low alarm until the
process first exceeds the alarm value. This suppresses false alarms during
system start-up. 

OUTPUT

ON CONDITION

RELAY

Energized

Annunciator

Readout

B/G

Max reading 

NET

Min Reading

T

Totalizer/Integrator 

Содержание MP30

Страница 1: ...uts implemented on Plug in cards The Plug in cards provide dual FORM C relays 5 A quad FORM A relays 3 A or either sinking or sourcing quad open collector logic outputs The setpoint alarms can be configured in modes to suit a variety of control and alarm requirements High and low absolute high and low deviation and band acting Balanced or unbalanced hysteresis On and off delay timers Auto reset or...

Страница 2: ...RS485 Isolation To Sensor User Input Commons 500 Vrms for 1 min Working Voltage 50 V Not Isolated from all other commons Data 7 8 bits Baud 300 to 19200 Parity no odd or even Bus Address selectable 0 to 99 Max 32 meters per line RS485 Transmit Delay Selectable for 2 to 50 msec or 50 to 100 msec RS485 17 ANALOG OUTPUT Types 0 to 20 mA 4 to 20 mA or 0 to 10 VDC Isolation To Sensor User Input Commons...

Страница 3: ... parameter and indexes to the next During Digital Tare this key stores the tare weight value keyed in KEY The Tare In Tare Out key both acquires and clears the Tare weight value The function of this key can be changed by setting the F1 function parameter For data entry this key is also used to increment a parameter value KEY The key Zero re zeros the Basic and Gross readout displays The function o...

Страница 4: ...as a basic millivoltmeter The basic indicator readout can then be post scaled to read out in weight units The indicator provides two ways in which to scale the display Key in Key in the input and display scaling points using known data Apply Apply the actual input value and key in the corresponding display value The indicator records the input value applied For processes that require linearity com...

Страница 5: ...ons Each time the batch function is executed the weight is added to the total and then zeroed PARAMETER LOCK MODE A user input can be used to lock the parameter list When the user input is active the indicator is in the protected parameter mode where it is only possible to access the setpoint values and the security code It is possible to lock the parameter list without using a user input as a pro...

Страница 6: ...deviation or band acting the effective trigger point is offset by SP1 In these modes SP1 acts as a master setpoint that the other setpoints track An alternate list of setpoint values can be stored and recalled as needed Select the alternate list with a function key or user input and program the setpoint values This allows for quick changeover for a new process condition Setpoint Alarm Hysteresis a...

Страница 7: ...eset Active Reset LIst Select main or alternate setpoint list r 1 Reset setpoint 1 r 2 Reset setpoint 2 USr 2 USr 1 User Input Function no no function pLOC Lock parameter list USr 3 rEL Acquire zero Settings 0 3 8 15 Active Reset Active Function Active Min Active Function Active Max Active Enable Active Enable Active Reset Active Total Active edge Batch Release from active Synchronize Active Hold ...

Страница 8: ...ale Factor sCfaC DESCRIPTION COMMENTS RANGE AND UNITS PARAMETER DISPLAY Tracking on reading band Tracking off reading 0 reading band PARAMETER MODULE 6 Setpoint Parameters DISPLAY PARAMETER RANGE AND UNITS DESCRIPTION COMMENTS act n Action for Setpoint ab HI absolute high balanced hysteresis SPSEL Select Setpoint Select setpoint for configuration Select operating action for setpoint See Figures fo...

Страница 9: ...sIN Output Source Assignment HI Max reading Select source of linear output signal If totalizer scaling limited to lower 5 digits tot Totalizer value 0 20 0 to 20 mA NEt Net weight b 6 Brutto Gross weight LO Min reading YES Enter sub menu t Tons NO None NO Disables parameter 0 False zero 6 Grams Print Options Line Print Units Suffix Opt UNIt tot Totalizer values 0oc Zero and ounces 6roSS Brutto Gro...

Страница 10: ...ers 5 tOt Totalizer Parameters 8 Out Analog Output Parameters 7 SrL Serial Communication Parameters DISPLAY PARAMETER USER SETTING FACTORY SETTING B 6 tArE BRUTTO GROSS DISPLAY TARE DISPLAY rEd Ent SP 3 SP 1 SETPOINT 3 SETPOINT 1 LOC LOC SP 4 SP 2 SETPOINT 4 SETPOINT 2 LOC LOC CodE SECURITY CODE 00000 OPEr HI t MOPERATING MODE MAX CAPTURE DELAY TIME b 6Nt 0000 0 LO t MIN CAPTURE DELAY TIME 0000 0 ...

Страница 11: ...rivers pens pencils etc to operate the keypad of the unit EMC GUIDELINES The indicator has been designed and tested to meet the requirements of EMC Directive 89 336 EEC However successful installations depend upon several factors These include cable routing termination of cable shields associated equipment etc Installation The indicator meets NEMA 4X IP65 requirements for indoor use when properly ...

Страница 12: ...med as a low acting alarm triggering on the Net liquid weight Each filling operation is started with a Tare Out Tare In sequence This opens the filling valve and when the target setpoint is reached the valve is shut off The valve remains off until a tare operation is repeated thereby restarting the filling operation The fourth output is connected in series with the fill output to disable filling d...

Отзывы: