SPECIFICATIONS

1. 

DISPLAY

: 5 digit, 0.56" (14.2 mm) red sunlight readable LEDs, (-19999 to

99999)

2. 

POWER

115/230 V Version (MP30/00)

AC Power: 85 to 250 VAC, 50/60 Hz, 15 VA
Isolation: 2300 Vrms for 1 min. to all inputs and outputs.

Working Voltage: 300 V.

24 V Version (MP30/01)

DC Power: 11 to 36 VDC, 11 W (derate operating temperature to 40°C if

operating <15 VDC and all three plug-in cards are installed)

AC Power: 24 VAC, ±10%, 50/60 Hz, 15 VA
Isolation: 500 Vrms for 1 min. to all inputs and outputs.

Working Voltage: 50 V.

3. 

ANNUNCIATORS

B/G

- Brutto/gross weight display

NET

- Net weight display

T

- Tare weight display

SP1

- setpoint alarm 1 is active

SP2

- setpoint alarm 2 is active

SP3

- setpoint alarm 3 is active

SP4

- setpoint alarm 4 is active

Units Label

- software controlled weight units label backlight

4. 

KEYPAD

: 5 keys

5. 

A/D CONVERTER

: 16 bit resolution

6. 

UPDATE RATES

:

A/D conversion rate: 20/readings sec
Step response: 200 msec. max. to within 99% of final readout value 

(digital filter and internal zero correction disabled)**
700 msec. max. (digital filter disabled, internal zero correction enabled)**

Display update rate: 1 to 20 updates/sec
Setpoint output on/off delay time: 0.0 to 3275.0 sec
Analog output update rate: 0.0 to 10.0 sec
Max./Min. capture delay time: 0.0 to 3275.0 sec

7. 

DISPLAY MESSAGES

:

“OLOL” - Appears when measurement e signal range.
“ULUL” - Appears when measurement exceeds - signal range
“. . . .” - Appears when display values display range.
“- . . .” - Appears when display values exceed - display range.

8.

CONNECTION TYPE

: 4 or 6 wire bridge 

9.

COMMON MODE RANGE (

w.r.t. input common): 0 to +5 VDC

Rejection: 80 db (DC to 120 Hz)

10.

SENSOR INPUTS

:

* After 20 minute warm-up. Accuracy is specified in two ways: Accuracy

over an 18 to 28

°

C and 10 to 75% RH environment; and accuracy over a 0 to

50

°

C and 0 to 85%RH (non-condensing environment). Accuracy over the 0 to

50

°

C range includes the temperature coefficient effect of the indicator. 

** The indicator periodically (every 12 seconds) imposes a 500 msec delay

to compensate for internal zero drift. If the delay affects applications where step
response is critical, it can be defeated. Set the display update to 20/sec to
disable. In this case, add a zero error of 0.2% FS to the 24 mV input range over
the 0 to 50

°

C span.

11. 

BRIDGE EXCITATION 

:

Jumper Selectable: 5 VDC @ 65 mA max., 

±

2%

10 VDC @ 125 mA max., 

±

2%

Temperature coefficient (ratio metric): 20 ppm/

°

C max.

12. 

LOW FREQUENCY NOISE REJECTION

:

Normal Mode: > 60 dB @ 50 or 60 Hz ±1%, digital filter off
Common Mode: >100 dB, DC to 120 Hz (w.r.t. earth)

13. 

USER INPUTS (Logic Level)

: Three software defined user inputs, jumper

selectable for sink/source logic

Max. Continuous Input: 30 VDC
Isolation To Sensor Input Common: Not isolated
User Inputs are isolated from all option card terminals.

14. 

TOTALIZER

:

Time Base: second, minute, hour, or day
Time Accuracy: 0.01% typical
Decimal Point: 0 to 0.0000

Scale Factor: 0.001 to 65.000
Low Signal Cut-out: -19,999 to 99,999
Total: 9 digits, display alternates between high order and low order readouts

15. 

CUSTOM LINEARIZATION

:

Data Point Pairs: Selectable from 2 to 12 
Display Range: -19,999 to 99,999
Decimal Point: 0 to 0.0000

16. 

SERIAL COMMUNICATIONS

: (RS232 or RS485)

Isolation To Sensor & User Input Commons: 500 Vrms for 1 min.

Working Voltage: 50 V.    Not Isolated from all other commons. 

Data: 7/8 bits
Baud: 300 to 19200
Parity: no, odd or even
Bus Address: selectable 0 to 99, Max. 32 meters per line (RS485)
Transmit Delay: Selectable for 2 to 50 msec or 50 to 100 msec (RS485)

17. 

ANALOG OUTPUT

:

Types: 0 to 20 mA, 4 to 20 mA or 0 to 10 VDC
Isolation To Sensor & User Input Commons: 500 Vrms for 1 min.

Working Voltage: 50 V.     Not Isolated from all other commons. 

Accuracy: 0.17% of FS (18 to 28°C); 0.4% of FS (0 to 50°C)
Resolution: 1/3500
Compliance: 10 VDC: 10 K

load min.

20 mA: 500 

load max.

18. 

SETPOINT OUTPUT

: Two types of field installable cards

Dual Relay Card

:

Type: Two FORM-C relays
Isolation To Sensor & User Input Commons: 2000 Vrms for 1 min.
Contact Rating: 

One Relay Energized: 5 amps @ 120/240 VAC or 28 VDC (resistive

load), 1/8 HP @120 VAC, inductive load

Total current with both relays energized not to exceed 5 amps

Life expectancy: 100K cycles min. at full load rating. External RC snubber

extends relay life for operation with inductive loads

Quad Relay Card

:

Type: Four FORM-A relays
Isolation To Sensor & User Input Commons: 2300 Vrms for 1 min.
Contact Rating: 

One Relay Energized: 3 amps @ 250 VAC or 30 VDC (resistive load), 

1/10 HP @120 VAC, inductive load

Total current with all four relays energized not to exceed 4 amps

Life Expectancy: 100K cycles min. at full load rating. External RC

snubber extends relay life for operation with inductive loads

19. 

ENVIRONMENTAL CONDITIONS

:

Operating Temperature Range: 0 to 50°C (0 to 45°C with all three plug-in

cards installed)

Storage Temperature Range: -40 to 60°C
Operating and Storage Humidity: 0 to 85% max. non-condensing
Altitude: Up to 2000 meters

20. 

CERTIFICATIONS AND COMPLIANCES

:

ELECTROMAGNETIC COMPATIBILITY

Notes:

1. Self-recoverable loss of performance during EMI disturbance at 10 V/m:

Measurement error exceeds unit specifications

For operation without loss of performance:

Mount unit in a metal enclosure (Buckeye SM7013-0 or equivalent)
Route power and I/O cables in metal conduit connected to earth ground.

21. 

CONNECTIONS

: High compression cage-clamp terminal block

Wire Strip Length:

0.35" (9 mm)

Wire Gauge Capacity

: One 14 AWG solid or Two 18 AWG

22. 

CONSTRUCTION

:This unit is rated for NEMA 4X/IP65 indoor use. IP20

Touch safe. Installation Category II, Pollution Degree 2. One piece
bezel/case. Flame resistant. Synthetic rubber keypad. Panel gasket and
mounting clip included.

23. 

WEIGHT

: 10.4 oz. (295 g)

Immunity to EN 50082-2

Electrostatic discharge

EN 61000-4-2 Level 3; 8 Kv air

Electromagnetic RF fields

EN 61000-4-3 Level 3; 10 V/m 

1

80 MHz - 1 GHz

Fast transients (burst)

EN 61000-4-4 Level 4; 2 Kv I/O

Level 3; 2 Kv power

RF conducted interference

EN 61000-4-6 Level 3; 10 V/rms 

Simulation of cordless telephones ENV 50204 Level 3; 10 V/m

900 MHz ± 5 MHz

150 KHz - 80 MHz

Emissions to EN 50081-2

RF interference

EN 55011

Enclosure class A
Power mains class A

200 Hz, 50% duty cycle

INPUT STATE

SOURCING INPUTS

22 K

pull-down

SINKING INPUTS

22 K

pull-up to +5 V

Active

V

IN

> 3.6 VDC

V

IN

< 0.9 VDC

Inactive

V

IN

< 0.9 VDC

V

IN

> 3.6 VDC

Input Range

Accuracy* 

(18 to 28C)

Accuracy* 

(0 to 50C)

Impedance

Resolution

±

24 mVDC

0.02% of 

r3 

µ

0.07% of

r4 

µ

V 100 Mohm

16 V

µ

V

±

240 mVDC

0.07% of

r40 

µ

V 100 Mohm

16 V

10 

µ

V

Max

Continuous

Overload

0.02% of 

r30 

µ

V

Содержание MP30

Страница 1: ...uts implemented on Plug in cards The Plug in cards provide dual FORM C relays 5 A quad FORM A relays 3 A or either sinking or sourcing quad open collector logic outputs The setpoint alarms can be configured in modes to suit a variety of control and alarm requirements High and low absolute high and low deviation and band acting Balanced or unbalanced hysteresis On and off delay timers Auto reset or...

Страница 2: ...RS485 Isolation To Sensor User Input Commons 500 Vrms for 1 min Working Voltage 50 V Not Isolated from all other commons Data 7 8 bits Baud 300 to 19200 Parity no odd or even Bus Address selectable 0 to 99 Max 32 meters per line RS485 Transmit Delay Selectable for 2 to 50 msec or 50 to 100 msec RS485 17 ANALOG OUTPUT Types 0 to 20 mA 4 to 20 mA or 0 to 10 VDC Isolation To Sensor User Input Commons...

Страница 3: ... parameter and indexes to the next During Digital Tare this key stores the tare weight value keyed in KEY The Tare In Tare Out key both acquires and clears the Tare weight value The function of this key can be changed by setting the F1 function parameter For data entry this key is also used to increment a parameter value KEY The key Zero re zeros the Basic and Gross readout displays The function o...

Страница 4: ...as a basic millivoltmeter The basic indicator readout can then be post scaled to read out in weight units The indicator provides two ways in which to scale the display Key in Key in the input and display scaling points using known data Apply Apply the actual input value and key in the corresponding display value The indicator records the input value applied For processes that require linearity com...

Страница 5: ...ons Each time the batch function is executed the weight is added to the total and then zeroed PARAMETER LOCK MODE A user input can be used to lock the parameter list When the user input is active the indicator is in the protected parameter mode where it is only possible to access the setpoint values and the security code It is possible to lock the parameter list without using a user input as a pro...

Страница 6: ...deviation or band acting the effective trigger point is offset by SP1 In these modes SP1 acts as a master setpoint that the other setpoints track An alternate list of setpoint values can be stored and recalled as needed Select the alternate list with a function key or user input and program the setpoint values This allows for quick changeover for a new process condition Setpoint Alarm Hysteresis a...

Страница 7: ...eset Active Reset LIst Select main or alternate setpoint list r 1 Reset setpoint 1 r 2 Reset setpoint 2 USr 2 USr 1 User Input Function no no function pLOC Lock parameter list USr 3 rEL Acquire zero Settings 0 3 8 15 Active Reset Active Function Active Min Active Function Active Max Active Enable Active Enable Active Reset Active Total Active edge Batch Release from active Synchronize Active Hold ...

Страница 8: ...ale Factor sCfaC DESCRIPTION COMMENTS RANGE AND UNITS PARAMETER DISPLAY Tracking on reading band Tracking off reading 0 reading band PARAMETER MODULE 6 Setpoint Parameters DISPLAY PARAMETER RANGE AND UNITS DESCRIPTION COMMENTS act n Action for Setpoint ab HI absolute high balanced hysteresis SPSEL Select Setpoint Select setpoint for configuration Select operating action for setpoint See Figures fo...

Страница 9: ...sIN Output Source Assignment HI Max reading Select source of linear output signal If totalizer scaling limited to lower 5 digits tot Totalizer value 0 20 0 to 20 mA NEt Net weight b 6 Brutto Gross weight LO Min reading YES Enter sub menu t Tons NO None NO Disables parameter 0 False zero 6 Grams Print Options Line Print Units Suffix Opt UNIt tot Totalizer values 0oc Zero and ounces 6roSS Brutto Gro...

Страница 10: ...ers 5 tOt Totalizer Parameters 8 Out Analog Output Parameters 7 SrL Serial Communication Parameters DISPLAY PARAMETER USER SETTING FACTORY SETTING B 6 tArE BRUTTO GROSS DISPLAY TARE DISPLAY rEd Ent SP 3 SP 1 SETPOINT 3 SETPOINT 1 LOC LOC SP 4 SP 2 SETPOINT 4 SETPOINT 2 LOC LOC CodE SECURITY CODE 00000 OPEr HI t MOPERATING MODE MAX CAPTURE DELAY TIME b 6Nt 0000 0 LO t MIN CAPTURE DELAY TIME 0000 0 ...

Страница 11: ...rivers pens pencils etc to operate the keypad of the unit EMC GUIDELINES The indicator has been designed and tested to meet the requirements of EMC Directive 89 336 EEC However successful installations depend upon several factors These include cable routing termination of cable shields associated equipment etc Installation The indicator meets NEMA 4X IP65 requirements for indoor use when properly ...

Страница 12: ...med as a low acting alarm triggering on the Net liquid weight Each filling operation is started with a Tare Out Tare In sequence This opens the filling valve and when the target setpoint is reached the valve is shut off The valve remains off until a tare operation is repeated thereby restarting the filling operation The fourth output is connected in series with the fill output to disable filling d...