Safety Summary
All safety related regulations, local codes and instructions that
appear in the manual or on equipment must be observed to ensure
personal safety and to prevent damage to either the instrument or
equipment connected to it. If equipment is used in a manner not
specified by the manufacturer, the protection provided by the
equipment may be impaired.
Do not use this unit to directly command motors, valves, or
other actuators not equipped with safeguards. To do so can be
potentially harmful to persons or equipment in the event of a fault
to the unit.
FRONT PANEL
JUMPER LINK FUNCTIONS
The indicator has three jumper links that must be set prior to operation.
These are:
•
Input Range Select
Select either
±
24 mVDC or
±
240 mVDC range. Select an input range
high enough to avoid input signal overload. Program the input range to
match jumper selection.
•
Bridge Excitation Select
Select either 5 VDC or 10 VDC bridge excitation.
•
User Input Source or Sink Select
FRONT
KEY FUNCTIONS
TARA KEY FUNCTION
The
TARA
key selects among the Gross, Net and Tare weight displays.
Corresponding panel enunciators illuminate to indicate the selected display. If
Tare weight = 0, the indicator prevents selection of Gross and Net weight
displays. Within the program menu, this key quits programming and returns to
display mode.
PAR KEY FUNCTION
The
PAR
key allows access to the configuration parameters. The function of
this key can be locked.
Within the program menu, this key stores the active parameter and indexes to
the next.
During Digital Tare, this key stores the tare weight value keyed in.
KEY
The
(Tare In/Tare Out) key both acquires and clears the Tare weight
value. The function of this key can be changed by setting the F1 function
parameter. For data entry, this key is also used to increment a parameter value.
KEY
The
key (Zero) re-zeros the Basic and Gross readout displays. The
function of this key can be changed by setting the F2 function parameter. For
data entry, this key is also used to decrement a parameter value.
Key
The
key (Print) causes a Print Ticket output. Prints occur regardless of
scale motion. The function of this key can be changed by setting the RST
function parameter. For data entry, this key is also used to scroll a parameter
value by x1000 factor when held with the Tare and Zero keys.
DISPLAYS
BASIC WEIGHT DISPLAY
The Basic Weight Display provides a direct indicator readout. The display
occurs when the Tare Weight =0.
BRUTTO/GROSS WEIGHT DISPLAY
This Display indicates the Brutto/Gross weight. It is only possible to select
this display mode when Tare > 0, otherwise it is locked out by the indicator.
“B/G” illuminates for this display mode.
NET WEIGHT DISPLAY
This Display indicates the Net weight. It is only possible to select this display
mode when Tare > 0, otherwise it is locked out by the indicator. If after the Tare
Weight is set > 0, the Net Display is selected by the indicator. “NET” illuminates
for this display mode.
TARE WEIGHT DISPLAY
This Display indicates the Tare weight. When this display is selected, it is
possible to directly key-in the Tare Weight value. “T” illuminates for this display
mode. It flashes during tare weight key-in entry.
!"!"!"!"!