background image

5

AUTOMATIC ZERO TRACKING

The indicator can be programmed to automatically compensate for zero drift.

Drift may be caused by changes in the transducers or electronics, or
accumulation of material on weight systems.

Auto-zero tracking operates whenever the readout remains within the

tracking band for a period of time equal to the tracking delay time. When these
conditions are met, the indicator re-zeroes the readout. After the re-zero
operation, the indicator resets and continues to auto-zero track.

The auto-zero tracking band should be set large enough to track normal zero

drift, but small enough to not interfere with small process inputs.

For filling operations, the fill rate must exceed the auto-zero tracking rate.

This avoids false tracking at the start of the filling operation.

Fill Rate  

tracking band

tracking time

Auto-zero tracking is disabled by setting the tracking band parameter = 0.

Note: The Automatic Zero operation is blocked when the total accumulated zero

exceeds the value of the Zero Acquisition Range. 

Note: The indicator does not have motion detection. Mechanical vibrations

coupled to the scale from other machinery may result in false Auto-Zero
operations.

DIGITAL FILTER

A unique adaptive input filter is used. Whenever the difference between one

reading and the next is less than the filter band value, the input is filtered. When
the difference exceeds the filter band value, the input is not filtered. This avoids
the usual compromise between using a relative high time constant for good
noise rejection and using a low time constant filter for quick step response.

FUNCTION KEYS

The Tare, Zero, and Print keys, in lieu of the standard function, can be

programmed for different functions, or no function at all. Additionally, the Tare
and Zero keys have second functions. The second function is activated after the
primary function if the key is held for 3 seconds.

MAX AND MIN READING DETECTION *

The indicator records the maximum (max) and minimum (min) process

inputs. Conditions such as valve activation, sudden change in material flow rate,
etc., can result in false peaks which are not reflective of the true maximum and
minimum of the process. In this case, Max and Min capture delay times can be
used to prevent the detection of false maximums and minimums.

TOTALIZER (integrator) *

The totalizer can be used to integrate a time-input product (such as flow). The

totalizer accumulates input readings according to the following relation:

Total(n) = Total(n-1) + [Input(n) * (scale factor/time base)]

n = nth  sample instant (seconds)
input = input reading (ignore decimal point position)
scale factor = 0.001 to 65.000
time base: sec = 1

min = 60

hour = 3600

day = 86,400

When the total exceeds 5 digits, the front panel annunciator “T” flashes. In

this case, the indicator continues to totalize up to a 9 digit value. The high order
4 digits and the low order 5 digits of the total are displayed alternately. The letter

h

” denotes the high order display.
The total decimal point position is independent of the input. A low cut value

disables totalization below a set input value. Alarms can be programmed to
activate when a set total is reached.

Alternately, the totalizer can also be used to accumulate batch weighing

operations. Each time the batch function is executed, the weight is added to the
total and then zeroed.

PARAMETER LOCK MODE

A user input can be used to lock the parameter list.When the user input is

active, the indicator is in the protected parameter mode, where it is only possible
to access the setpoint values and the security code. 

It is possible to lock the parameter list without using a user input as a program

lock function. In this case, set the security code to a non-zero value. With a non-
zero security value set, press the PAR key to view the programmed setpoint
values. The security code requires a matching “key” value to gain access to the
full parameter list.

BASIC OPERATION

TARE WEIGHT REGISTER

The indicator has a single Tare Weight register. There are two modes in which

Tare data is acquired: Pushbutton Tare and Digital Tare.

PUSHBUTTON TARE

Pushbutton Tare is activated by a single key press of the Tare In/Tare Out

Key.

Acquire Tare (Tare In):

Tare weight must be cleared, and Basic Weight Displayed.
Place Tare weight on scale.
Allow readout to stabilize.
Press key.
Indicator acquires the Tare weight with these actions: 
If new Tare weight value >0, indicator displays “

'()*

” and selects Net

Weight display.

If new Tare weight < or = 0, indicator rejects Tare weight and displays

&+'()

”.

Clear tare (Tare Out):

Select Gross or Net Weight display.
Tare weight can be either on or off scale.
Press

to clear Tare weight

Indicator selects Basic Weight display.

DIGITAL TARE 

Digital Tare allows a direct entry of  the Tare Weight.

Enter Digital Tare:

Select the Tare Weight display. 
Enter the Tare Weight using the Tare, Zero and Print keys. In this mode,

they take on data entry functions. Use the Tare key as an increment key
and the Zero key as a decrement key. 

The “T” annunciator flashes to indicate Tare Weight entry.
Press the 

PAR

key to register the Tare weight value. The indicator selects

either Net Weight display (new Tare Weight >0) or Basic weight display
(new Tare weight = 0). 

Press the 

TARA

key to abort the Tare entry mode. 

PUSHBUTTON ZERO 

Pushbutton Zero is activated by a single key press of the Zero Key.

To Zero:

Select Basic or Gross Weight display.
Allow readout to stabilize
Press

Key to re-zero the Gross Weight. Meter displays “

)*,*'

”.

Note: This operation is blocked when the total accumulated zero exceeds

the value of the Zero Acquisition Range.  

PRINT TICKET

A ticket print is activated by a single key press of the Print Key.

To Print:

Place weight on scale and allow readout to stabilize.
Press

key.

(Print format is programmable in Module 

-.$)/

.)

AUTOMATIC AND MANUAL ZERO ACQUISITION

There are two means by which the indicator keeps an accurate zero: Automatic

and Manual. The indicator saves the new zero state in an identical way in which
the Tare Weight is saved. See Tare Weight Entry for details of operation.

AUTOMATIC ZERO ACQUISITION

In Automatic mode, the indicator automatically re-zeros when the Gross

weight falls within a specified amount of scale divisions of zero (given by the
Autozero track band parameter) and remains within this band for a specified
amount of time (given by the Autozero track time). The indicator continues to
track zero in this fashion until the total accumulated zero offset exceeds the Zero
Acquisition Range. When exceeded, Autozero tracking is disabled. 

Note: The indicator does not incorporate a means to detect stability. As a result,

scale vibration about zero may cause a false Auto-Zero operation.

Содержание MP30

Страница 1: ...uts implemented on Plug in cards The Plug in cards provide dual FORM C relays 5 A quad FORM A relays 3 A or either sinking or sourcing quad open collector logic outputs The setpoint alarms can be configured in modes to suit a variety of control and alarm requirements High and low absolute high and low deviation and band acting Balanced or unbalanced hysteresis On and off delay timers Auto reset or...

Страница 2: ...RS485 Isolation To Sensor User Input Commons 500 Vrms for 1 min Working Voltage 50 V Not Isolated from all other commons Data 7 8 bits Baud 300 to 19200 Parity no odd or even Bus Address selectable 0 to 99 Max 32 meters per line RS485 Transmit Delay Selectable for 2 to 50 msec or 50 to 100 msec RS485 17 ANALOG OUTPUT Types 0 to 20 mA 4 to 20 mA or 0 to 10 VDC Isolation To Sensor User Input Commons...

Страница 3: ... parameter and indexes to the next During Digital Tare this key stores the tare weight value keyed in KEY The Tare In Tare Out key both acquires and clears the Tare weight value The function of this key can be changed by setting the F1 function parameter For data entry this key is also used to increment a parameter value KEY The key Zero re zeros the Basic and Gross readout displays The function o...

Страница 4: ...as a basic millivoltmeter The basic indicator readout can then be post scaled to read out in weight units The indicator provides two ways in which to scale the display Key in Key in the input and display scaling points using known data Apply Apply the actual input value and key in the corresponding display value The indicator records the input value applied For processes that require linearity com...

Страница 5: ...ons Each time the batch function is executed the weight is added to the total and then zeroed PARAMETER LOCK MODE A user input can be used to lock the parameter list When the user input is active the indicator is in the protected parameter mode where it is only possible to access the setpoint values and the security code It is possible to lock the parameter list without using a user input as a pro...

Страница 6: ...deviation or band acting the effective trigger point is offset by SP1 In these modes SP1 acts as a master setpoint that the other setpoints track An alternate list of setpoint values can be stored and recalled as needed Select the alternate list with a function key or user input and program the setpoint values This allows for quick changeover for a new process condition Setpoint Alarm Hysteresis a...

Страница 7: ...eset Active Reset LIst Select main or alternate setpoint list r 1 Reset setpoint 1 r 2 Reset setpoint 2 USr 2 USr 1 User Input Function no no function pLOC Lock parameter list USr 3 rEL Acquire zero Settings 0 3 8 15 Active Reset Active Function Active Min Active Function Active Max Active Enable Active Enable Active Reset Active Total Active edge Batch Release from active Synchronize Active Hold ...

Страница 8: ...ale Factor sCfaC DESCRIPTION COMMENTS RANGE AND UNITS PARAMETER DISPLAY Tracking on reading band Tracking off reading 0 reading band PARAMETER MODULE 6 Setpoint Parameters DISPLAY PARAMETER RANGE AND UNITS DESCRIPTION COMMENTS act n Action for Setpoint ab HI absolute high balanced hysteresis SPSEL Select Setpoint Select setpoint for configuration Select operating action for setpoint See Figures fo...

Страница 9: ...sIN Output Source Assignment HI Max reading Select source of linear output signal If totalizer scaling limited to lower 5 digits tot Totalizer value 0 20 0 to 20 mA NEt Net weight b 6 Brutto Gross weight LO Min reading YES Enter sub menu t Tons NO None NO Disables parameter 0 False zero 6 Grams Print Options Line Print Units Suffix Opt UNIt tot Totalizer values 0oc Zero and ounces 6roSS Brutto Gro...

Страница 10: ...ers 5 tOt Totalizer Parameters 8 Out Analog Output Parameters 7 SrL Serial Communication Parameters DISPLAY PARAMETER USER SETTING FACTORY SETTING B 6 tArE BRUTTO GROSS DISPLAY TARE DISPLAY rEd Ent SP 3 SP 1 SETPOINT 3 SETPOINT 1 LOC LOC SP 4 SP 2 SETPOINT 4 SETPOINT 2 LOC LOC CodE SECURITY CODE 00000 OPEr HI t MOPERATING MODE MAX CAPTURE DELAY TIME b 6Nt 0000 0 LO t MIN CAPTURE DELAY TIME 0000 0 ...

Страница 11: ...rivers pens pencils etc to operate the keypad of the unit EMC GUIDELINES The indicator has been designed and tested to meet the requirements of EMC Directive 89 336 EEC However successful installations depend upon several factors These include cable routing termination of cable shields associated equipment etc Installation The indicator meets NEMA 4X IP65 requirements for indoor use when properly ...

Страница 12: ...med as a low acting alarm triggering on the Net liquid weight Each filling operation is started with a Tare Out Tare In sequence This opens the filling valve and when the target setpoint is reached the valve is shut off The valve remains off until a tare operation is repeated thereby restarting the filling operation The fourth output is connected in series with the fill output to disable filling d...

Отзывы: