4

6-Wire Bridge Input With User Input Sourcing Logic Connection

4-Wire Bridge Input With User Input Sinking Logic Connection

BASIC CONNECTIONS

Sensor Input Common is NOT isolated

from User Input Common.

* For 6-wire load cell connections, the +SEN and -SEN leads may

be tied in parallel with the +EXC and -EXC respectively, or the sense
leads may be left unconnected.

WARNING

: Disconnect power to the unit before installing plug-in

cards. To install a card, depress the side latches at the finger holds
and remove the unit from the rear of the case. Insert the card into the
appropriate slot and re-install the unit.

INPUTS

The indicator has two input ranges: 

±

24 mV DC or 

±

240 mV DC. Set the

internal jumper link to select the desired range. Select the matching range in
programming. Select a range that is high enough to accomodate the bridge
output to avoid overload. To allow use of the higher sensitivity 24 mV input
range, use 5 V excitation with high output (3 mV/V) bridges. 5 V excitation also
reduces bridge power compared to 10 V excitation.

A maximum of four 350 ohm load cells can be driven by the internal bridge

excitation voltage.

DEADLOAD COMPENSATION

In some cases, the combined deadload and liveload output may exceed the

range of the 24 mV input. To use this range, the output of the bridge can be
offset a small amount by applying a fixed resistor across one arm of the bridge.
This shifts the electrical output of the bridge to within the operating range of the
indicator. A 100 K fixed resistor shifts the bridge output approximately -10 mV
(350 ohm bridge, 10 V excitation).

Apply the resistor SIG to signal common. Use a metal film resistor

with a low temperature coefficient of resistance.

SCALING

The indicator has been factory calibrated on both ranges as a basic

millivoltmeter. The basic indicator readout can then be post-scaled to read out
in weight units. The indicator provides two ways in which to scale the display:

Key-in: Key in the input and display scaling points using known data.
Apply: Apply the actual input value and key in the corresponding display

value. The indicator records the input value applied.

For processes that require linearity compensation, up to 12 scaling points can

be used for correction. The indicator interprets the last 

#$%&

parameter as the

full scale capacity of the system. This determines the number of leading zeros
for display. The scaling range is extended up to five digits of resolution with
selectable scale division units. 6-digit readouts are afforded by installing a
dummy zero unit label.

READ RATE

The display readout, max/min reading, setpoint outputs and analog update

have independent update rates. The fastest update rate (20 readings/sec) ensures
detection of fast events (such as burst testing, pressure, and peak weight, etc.)

MP30/00: 85-250 VAC
MP30/01: 24 V AC/DC

MP30/00: 85-250 VAC
MP30/01: 24 V AC/DC

WEIGHT UNITS OVERLAY

The indicator has a backlighted units indicator that can be customized to the

application. The backlight is turned on by programming the “

b-LIt

” parameter.

Overlays are supplied with the indicator. To install an overlay, remove the unit
from the case. Select the label and apply it to the label frame, noting that the
label must be aligned accurately. Install the label frame to the display board in
the alignment holes located on the right side of the display.

PLUG-IN CARDS

The MP30 indicator has provisions for inclusion of three plug-in option

cards. Connection data is provided with each card. These cards are:

Setpoint Outputs
Analog Outputs
Communication Option

The plug-in cards can be used in any combination, however, it is only

possible to use one type of card from each category. Cards can be installed
initially, or at a later date as system needs arise.

Содержание MP30

Страница 1: ...uts implemented on Plug in cards The Plug in cards provide dual FORM C relays 5 A quad FORM A relays 3 A or either sinking or sourcing quad open collector logic outputs The setpoint alarms can be configured in modes to suit a variety of control and alarm requirements High and low absolute high and low deviation and band acting Balanced or unbalanced hysteresis On and off delay timers Auto reset or...

Страница 2: ...RS485 Isolation To Sensor User Input Commons 500 Vrms for 1 min Working Voltage 50 V Not Isolated from all other commons Data 7 8 bits Baud 300 to 19200 Parity no odd or even Bus Address selectable 0 to 99 Max 32 meters per line RS485 Transmit Delay Selectable for 2 to 50 msec or 50 to 100 msec RS485 17 ANALOG OUTPUT Types 0 to 20 mA 4 to 20 mA or 0 to 10 VDC Isolation To Sensor User Input Commons...

Страница 3: ... parameter and indexes to the next During Digital Tare this key stores the tare weight value keyed in KEY The Tare In Tare Out key both acquires and clears the Tare weight value The function of this key can be changed by setting the F1 function parameter For data entry this key is also used to increment a parameter value KEY The key Zero re zeros the Basic and Gross readout displays The function o...

Страница 4: ...as a basic millivoltmeter The basic indicator readout can then be post scaled to read out in weight units The indicator provides two ways in which to scale the display Key in Key in the input and display scaling points using known data Apply Apply the actual input value and key in the corresponding display value The indicator records the input value applied For processes that require linearity com...

Страница 5: ...ons Each time the batch function is executed the weight is added to the total and then zeroed PARAMETER LOCK MODE A user input can be used to lock the parameter list When the user input is active the indicator is in the protected parameter mode where it is only possible to access the setpoint values and the security code It is possible to lock the parameter list without using a user input as a pro...

Страница 6: ...deviation or band acting the effective trigger point is offset by SP1 In these modes SP1 acts as a master setpoint that the other setpoints track An alternate list of setpoint values can be stored and recalled as needed Select the alternate list with a function key or user input and program the setpoint values This allows for quick changeover for a new process condition Setpoint Alarm Hysteresis a...

Страница 7: ...eset Active Reset LIst Select main or alternate setpoint list r 1 Reset setpoint 1 r 2 Reset setpoint 2 USr 2 USr 1 User Input Function no no function pLOC Lock parameter list USr 3 rEL Acquire zero Settings 0 3 8 15 Active Reset Active Function Active Min Active Function Active Max Active Enable Active Enable Active Reset Active Total Active edge Batch Release from active Synchronize Active Hold ...

Страница 8: ...ale Factor sCfaC DESCRIPTION COMMENTS RANGE AND UNITS PARAMETER DISPLAY Tracking on reading band Tracking off reading 0 reading band PARAMETER MODULE 6 Setpoint Parameters DISPLAY PARAMETER RANGE AND UNITS DESCRIPTION COMMENTS act n Action for Setpoint ab HI absolute high balanced hysteresis SPSEL Select Setpoint Select setpoint for configuration Select operating action for setpoint See Figures fo...

Страница 9: ...sIN Output Source Assignment HI Max reading Select source of linear output signal If totalizer scaling limited to lower 5 digits tot Totalizer value 0 20 0 to 20 mA NEt Net weight b 6 Brutto Gross weight LO Min reading YES Enter sub menu t Tons NO None NO Disables parameter 0 False zero 6 Grams Print Options Line Print Units Suffix Opt UNIt tot Totalizer values 0oc Zero and ounces 6roSS Brutto Gro...

Страница 10: ...ers 5 tOt Totalizer Parameters 8 Out Analog Output Parameters 7 SrL Serial Communication Parameters DISPLAY PARAMETER USER SETTING FACTORY SETTING B 6 tArE BRUTTO GROSS DISPLAY TARE DISPLAY rEd Ent SP 3 SP 1 SETPOINT 3 SETPOINT 1 LOC LOC SP 4 SP 2 SETPOINT 4 SETPOINT 2 LOC LOC CodE SECURITY CODE 00000 OPEr HI t MOPERATING MODE MAX CAPTURE DELAY TIME b 6Nt 0000 0 LO t MIN CAPTURE DELAY TIME 0000 0 ...

Страница 11: ...rivers pens pencils etc to operate the keypad of the unit EMC GUIDELINES The indicator has been designed and tested to meet the requirements of EMC Directive 89 336 EEC However successful installations depend upon several factors These include cable routing termination of cable shields associated equipment etc Installation The indicator meets NEMA 4X IP65 requirements for indoor use when properly ...

Страница 12: ...med as a low acting alarm triggering on the Net liquid weight Each filling operation is started with a Tare Out Tare In sequence This opens the filling valve and when the target setpoint is reached the valve is shut off The valve remains off until a tare operation is repeated thereby restarting the filling operation The fourth output is connected in series with the fill output to disable filling d...