40
Service Manual Combics
Error description
After completing the internal test, the individual test items are displayed with their respective results.
A successful test is confirmed with the message "Test OK." An attempt is always made to test options such as E5 or A10.
If these options are not available, then the test results would say NOT FOUND. However, if the corresponding option is installed and
"NOT FOUND" appears, then this means the test was unsuccessful.
Several different error messages are displayed depending on the type of error.
If a COM interface is equipped with the YCC02-D09M6 cable, this triggers an error message because this cable is not available via the
DTR line that is checked by the test program.
This error message can then be ignored.
The following error messages are possible:
Operating Design | Error Description Test Combics | 1 | 2 | 3
Test
Message
Meaning
Possible error sources
None, but
Back light is always dark
-Display unit/Cable defective
-Cable not inserted correctly into the digital
PCB
-Digital PCB defective
Back light is always light
-Display unit/Cable defective
-Digital PCB defective
Display
not all segments are displayed
-Display unit defective
LED
No message
LEDs do not light as described in
Fehler! Verweisquelle konnte nicht
gefunden werden.
-Cable not connected or connected
incorrectly to display unit
-Cable to display unit defective
-LED module in display unit defective
-Digital PCB defective
KEY-PRESS Key was not recognized -Key not pressed in time
-Display unit/Cable defective
-Cable not inserted correctly into the digital
PCB
-Digital PCB defective
KEYBOARD
IRQ
Keypad interrupt not
detected
-Display unit/Cable defective
-Cable not inserted correctly into the digital
PCB
-Digital PCB defective
SRAM
ACCESS
Access to SRAM failed -Digital PCB defective
ACCESS
Failed access to
EEPROM
-Digital PCB defective
EEPROM
I2C
Protocol error
-Digital PCB defective
ACCESS
Failed access to alibi
memory
SPI
Protocol error
FLASH
No
symbol
If E5 is available:
Alibi memory not
detected
-Digital PCB defective
-Alibi memory defective
-Alibi memory not inserted correctly
Battery Battery
symbol
does not
appear
If L9 is available:
No communication
between digital PCB and
battery controller
-Connection cable between digital PCB and
battery controller not inserted or defective
-Battery controller defective
-Digital PCB defective
SCL2-ADU NOT FOUND
If A10 is available:
No communication
between digital PCB and
10.000e ADC
-Option A7 installed (not allowed with A10)
-Connection cable between digital PCB and
ADC not inserted or defective
-ADC defective
-Digital PCB defective
SCL1-RS232 LOOPBACK
If A6 is available:
No data connection via
the IS interface
-Option A8 or A9 installed (not allowed with
A6)
-Test adapter missing
-Connection cable between digital PCB and
interface not inserted or defective
-A6 interface defective
-Digital PCB defective
SCL1-RS485 LOOPBACK
If A6 is available:
No data connection via
the IS interface
-Option A8 or A9 installed (not allowed with
A6)
-Connection cable between digital PCB and
Interface not inserted or defective
-A6 interface defective
-Digital PCB defective
SCL2-RS232 LOOPBACK
If A7 is available:
No data connection via
the IS interface
-Option A10 (not allowed with A7)
-Test adapter missing
-Connection cable between digital PCB and
interface not inserted or defective
-A7 interface defective
-Digital PCB defective
Содержание CAIS3