Service Manual Combics
y
Select the ADC Configuration menu item and confirm:
t
Press the
k
key several times,
until "
ADC-CON
" appears, press the
)
key.
y
Select whether a standard configuration "
STAND
" or a verifiable configuration
"
VERIF.
" should be carried out.
Important!
After carrying out the required adjustments,
the adjustment function
that the customer should carry out must be set.
Operating Design | Service Mode Combics 1 | 2 ADC configuration
S
PARAM.!
K
ADU KON
)
S
y
Standard configuration "
STANDRD
.
"
Range
selection:
"
RANGE
"
Single-range mode
“
SINGLE
"
..
"D"
"MAX."
Multi-interval mode
“
MULT.INT
"
"D"
"RANGE
1"
"RANGE
2"
"RANGE
3"
"MAX."
Multiple-range mode
“
MULT.RNG
"
"D"
"RANGE
1"
"RANGE
2"
"RANGE
3"
"MAX."
Unit
(select the unit)
"
UNIT
"
e.g.: kilogram or ton
“
KG
"
“
T
"
Memory:
"
SAVE
"
Save
“
NO
"
Save
“
Yes
"