![Minebea Intec CAIS3 Скачать руководство пользователя страница 76](http://html.mh-extra.com/html/minebea-intec/cais3/cais3_service-manual_1801133076.webp)
75
Service Manual Combics
4-20mA / 0-10V interface, Option (A9)
Slot for UNICOM 4-20mA /0-10V Interface.
The operating mode setting is carried out via switches SW1-1 and SW1-2.
The PCB factory setting is 4-20mA operation. The switch must be reset to enable the card to
operate via 0-10V.
Important!
h
Switch settings must be made when replacing the PCB.
The following lists the configuration options:
Connection Function
1
I_out(+)
2
I_in(-)
3
V_out(+)
4
V_in (-)
5
GND galvanic isolation
6
GND galvanic isolation
SW1-1 SW1-2
Mode
Open
Open
4-20mA
Closed
Open
0-10V
Open
Closed
0-20mA
Closed
Closed
0-24mA
Interface Description | Unicom PCBs