19
Service Manual Combics
Operating Design | Service Mhhode Combics3
Configuring the A/D converter Combics | 1 | 2
Open the menu access switch
t
Remove the cap that covers the menu access switch on the back of the indicator
t
Slide the switch to the left (= "open position")
Enter the service password as described above (202122), confirm with the
)
key and
scroll back using the
(
key.
y
The device is now in Service mode. The letter S is displayed on the right in the Setup
menu header.
t
Select the "
SETUP
" menu item:
Press the
)
key.
t
Select weighing platform and confirm:
Press the
)
key.
t
Select the "
INTERNAL
" menu item:
t
Press the
)
key
setup
)
WP 1
)
INTERNAL
)
S
S
S
o