37
Service Manual Combics
Important!
h
However, this does not apply to the B9 (Ethernet), B1 (profibus) and A3 (RS-485)
options.
These options do not require a test adapter.
The test adapter must be supplied with a mains adapter (15V=) for option A5.
Test Implementation
t
Switch Combics to Standby mode.
t
Insert the test adapter into the interfaces (e.g. into COM 1).
t
Press and hold the
k
key.
t
Press the
e
key to turn the device back on.
y
You will here a brief beep tone three times (test has started).
t
Release the
k
key.
y
The display shows the version number of the firmware.
t
"DISPLAY TEST" appears on the display.
t
The display back lighting turns off and then on again.
y
All segments in the display are tested with a checkerboard pattern.
y
Carry out the visual inspection.
Operating Design | Test Program Combics 3