C - 2
Defibrillator/Monitor Operator’s Manual
If the device is operated within the electromagnetic environment listed in Table
Guidance and Declaration —
Electromagnetic Immunity
, the equipment will remain safe and provide the following essential performance:
HR accuracy, Resp accuracy, SpO
2
accuracy, PR accuracy, NIBP accuracy, CO
2
accuracy, Pacing rate accuracy,
Pacing output accuracy, energy accuracy, CPR function, alarm, data stored, user’s interface function.
Guidance and Declaration - Electromagnetic Immunity
The equipment is suitable for use in the electromagnetic environment specified below. The customer or the user of the
equipment should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD) IEC
61000-4-2
±8 kV contact
±15kV air
±8 kV contact
±15kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
(length greater than 3 m)
±2 kV for power supply
lines
±1 kV for input/output
lines
(length greater than 3 m)
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Voltage dips and
Voltage interruptions
IEC 61000-4-11
0% U
T
for 0,5 cycle
0% U
T
for 1 cycle and 70%
U
T
for 25/30 cycles
0% U
T
for 250/300 cycle
0% U
T
for 0,5 cycle
0% U
T
for 1 cycle and 70%
U
T
for 25/30 cycles
0% U
T
for 250/300 cycle
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of our product requires continued
operation during power mains
interruptions, it is recommended
that our product be powered
from an uninterruptible power
supply or a battery.
RATED power
frequency magnetic
fields IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
U
T
is the A.C. mains voltage prior to application of the test level.
Содержание BeneHeart D2
Страница 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Страница 2: ......
Страница 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 204: ......
Страница 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Страница 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 207: ......
Страница 208: ...P N 046 010599 00 9 0...