Defibrillator/Monitor Operator’s Manual
E - 3
SpO2
SpO2 Sensor Off
L*
B
The SpO
2
sensor has become detached from the patient
or the module, or there is a fault with the SpO
2
sensor, or
an unspecified SpO
2
sensor has been used. Check the
sensor application site and the sensor type, and make
sure the sensor is not damaged. Reconnect the sensor or
use a new sensor.
SpO2 Sensor Fault
L
C
SpO2 No Sensor
L
B
SpO2 Unknow Sensor
L
C
SpO2 Sensor
Incompatible
L
C
SpO2 Too Much Light
L
C
There is too much light on the SpO
2
sensor. Move the
sensor to a place with lower level of ambient light or cover
the sensor to minimize the ambient light.
SpO2 Low Signal
L
C
The SpO
2
signal is too low or too weak. Check the
patient’s condition and change the sensor application
site. If the error persists, replace the sensor.
SpO2 Weak Signal
L
C
SpO2 Weak Pulse
L
C
SpO2 Non-Pulsatile
L
C
SpO2 Low Perf
L
B
SpO2 Interference
L
C
The SpO
2
signal has been interfered. Check for any
possible sources of signal noise form the area around the
sensor, and check the patient for excessive motion.
SpO2 Board Fault
L
C
There is a problem with the SpO
2
measurement board. Do
not use the module and contact your service personnel.
NIBP
NIBP Loose Cuff
L
A
The NIBP cuff is not properly connected, or there is a leak
in the airway.
NIBP Air Leak
L
A
NIBP Pneumatic Leak
L
A
Check the NIBP cuff and pump for leakages.
NIBP Cuff Type Wrong
L
A
The cuff type applied mismatches the patient category.
Verify the patient category and replace the cuff.
NIBP Air Press. Err
L
A
An error occurred to the air pressure. Verify that the
equipment application site meets the environmental
requirements and check if there is any source that affects
the air pressure.
NIBP Weak Signal
L
A
The patient’s pulse is weak or the cuff is loose. Check the
patient’s condition and change the cuff application site. If
the problem persists, change the cuff.
NIBP Sig. Saturated
L
A
The NIBP signal is saturated due to excess motion or other
sources.
NIBP Overrange
L
A
The patient’s NIBP value may be beyond the specified
measurement range.
NIBP Excessive Motion
L
A
Check the patient’s condition and reduce the patient
motion.
NIBP Equip Err
H
A
An error occurred during NIBP measurement and
therefore the equipment cannot perform analysis
correctly. Check the patient’s condition and NIBP
connections, or replace the cuff.
NIBP Time Out
L
A
NIBP Measure Failed
L
A
NIBP Reset For Err
L
A
An illegal reset occurred during NIBP measurement.
Check if the airway is occluded.
Measurement
Alarm Message
L
I
Cause and solution
Содержание BeneHeart D2
Страница 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Страница 2: ......
Страница 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 204: ......
Страница 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Страница 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 207: ......
Страница 208: ...P N 046 010599 00 9 0...