A - 14
Defibrillator/Monitor Operator’s Manual
A.7
Recorder Specifications
A.8
Alarm Specifications
A.9
Data Storage
Method
High-resolution thermal dot array
Number of waveforms
3 at maximum
Paper speed
6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s. Error no more than
±
5%
Paper width
50 mm
Grid lines
The operator can choose to print grid lines or not
Auto record
Charge events, shock events, marked events, auto test report, parameter
alarms, ARR alarms, if configured on
Alarm Levels
High, medium, low level alarms, complying with IEC60601-1-8
Alarm Categories
Physiological alarms, technical alarms; Latched alarms and unlatched
alarms.
Alarm lamp
Independent alarm LED
Parameter alarm setting
Alarm properties of all available parameters can be set simultaneously in the
Para. Alarm menu
Auto alarm limits
Parameter alarm limits can be automatically adjusted according to currently
measured vital signs
Internal storage
1G Bytes
Marking events
16 types of events, user customized
Events
At least 1000 events for each patient.
Waveforms
At least 24 hours of consecutive ECG waveform
Voice recording
(available in the AED mode)
At least 180 minutes in total, more than 60 minutes for each patient
Data export
Data can be export to a PC through a USB flash memory
Tabular trends
Up to 72 hours for all measured parameters with the resolution no less than
1 minute.
Patient archives
Up to 100
Содержание BeneHeart D2
Страница 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Страница 2: ......
Страница 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 204: ......
Страница 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Страница 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 207: ......
Страница 208: ...P N 046 010599 00 9 0...