13 - 4
Defibrillator/Monitor Operator’s Manual
WARNING
•
Continuous non-invasive blood pressure measurements may cause purpura, ischemia and
neuropathy in the limb with the cuff. Inspect the application site regularly to ensure skin quality and
inspect the extremity of the cuffed limb for normal color, warmth and sensitivity. If any abnormity
occurs, move the cuff to another site or stop the blood pressure measurements immediately.
13.6.6
Starting a STAT Measurement
1.
Select the NIBP parameter area to enter the [
NIBP Setup
] menu.
2.
Select [
NIBP STAT
]. The STAT mode initiates a 5-minute continuous, automatic NIBP measurement.
13.7
Setting Initial Cuff Inflation Pressure
You can set the initial cuff inflation pressure manually.
1.
Select the NIBP parameter area to enter the [
NIBP Setup
] menu.
2.
Select [
Initial Pressure
] and then select the appropriate setting. The cuff is inflated accordingly at the next
NIBP measurement.
NOTE
•
For known hypertensive patients, you need to set initial cuff pressure to a higher value to reduce the
measurement time.
•
Setting initial cuff inflation pressure is disabled during NIBP measurement.
•
The initial cuff inflation pressure is restored to the default setting if NIBP module has been reset or
patient category has been changed.
13.8
Setting Pressure Unit
You can set [
Press. Unit
] to change NIBP unit only by selecting [
NIBP Setup
] from the Configuration Main menu.
13.9
NIBP Troubleshooting
.
Содержание BeneHeart D2
Страница 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Страница 2: ......
Страница 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 204: ......
Страница 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Страница 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 207: ......
Страница 208: ...P N 046 010599 00 9 0...