A - 4
Defibrillator/Monitor Operator’s Manual
NOTE
•
The equipment startup time in the fast startup mode is less than 2s.
Impedance
Selected energy
25Ω
50Ω
75Ω
100Ω
125Ω
150Ω
175Ω
Accuracy
1 J
1
1
1
0.9
0.9
0.9
0.8
±2J
2 J
2
2
2
1.9
1.8
1.7
1.6
±2J
3 J
2.9
3
2.9
2.8
2.7
2.6
2.4
±2J
4 J
3.9
4
3.9
3.7
3.6
3.4
3.2
±2J
5 J
4.9
5
4.9
4.7
4.5
4.3
4.1
±2J
6 J
5.8
6
5.8
5.6
5.3
5.1
4.9
±2J
7 J
6.8
7
6.8
6.6
6.3
6
5.7
±2J
8 J
7.8
8
7.8
7.4
7.1
6.8
6.5
±2J
9 J
8.8
9
8.8
8.4
8
7.7
7.3
±2J
10 J
9.7
10
9.7
9.3
8.9
8.5
8.1
±2J
15 J
15
15
15
14
13
13
12
±15%
20 J
20
20
20
19
18
17
16
±15%
30 J
29
30
29
28
27
25
24
±15%
50 J
49
50
49
47
45
43
41
±15%
70 J
68
70
68
65
62
60
57
±15%
100 J
97
100
97
93
89
85
81
±15%
150 J
146
150
146
140
134
128
122
±15%
170 J
166
170
166
159
151
145
138
±15%
200 J
195
200
195
187
178
170
163
±15%
300 J
292
300
292
280
267
255
244
±15%
360 J
351
360
350
336
321
306
293
±15%
Charge time (Note: at 20 ±5 °C of ambient temperature)
Manual Defib
AED
Charge time
From initial
power on (from
cold start) to
charge done
From initial
power on (from
fast startup
mode) to charge
done
From initiation
of rhythm
analysis to
charge done
From initial
power on (from
cold start) to
charge done
From initial
power on (from
fast startup
mode) to charge
done
200J
360J
200J
360J
200J
360J
200J
360J
200J
360J
200J
360J
With a new, fully
charged battery
<3 s
<7 s
<11 s
<14 s
<6 s
<10 s
<10 s
<12 s
<21 s
<26 s
<13 s
<15 s
With a new, fully
charged battery,
depleted by 15 360 J
discharges
<4 s
<8 s
<12 s
<15 s
<7 s
<11 s
<11 s
<13 s
<23 s
<27 s
<14 s
<16 s
With 90% to 100%
rated mains voltage
<4 s
<7 s
<11 s
<14 s
<7 s
<10 s
<11 s
<12 s
<22 s
<24 s
<14 s
<15 s
Содержание BeneHeart D2
Страница 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Страница 2: ......
Страница 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 204: ......
Страница 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Страница 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 207: ......
Страница 208: ...P N 046 010599 00 9 0...