Defibrillator/Monitor Operator’s Manual
E - 5
Power board
Power Board Selftest Err
H
C
An error occurred to the system power supply. Restart the
equipment.
Power Board Volt Err
L
C
Low Battery
S
C
Change battery or connect the equipment to the AC
power source to charge the batteries.
No Battery
L
C
Battery is not installed. Install the battery.
Battery Depleted!
System will shut down
immediately. Connect
to AC mains or replace
battery.
S
C
Connect the equipment to AC mains.
Battery Err
H
C
There is a problem with the batteries. Check the batteries
for damage; verify that correct batteries are used. Replace
the batteries if necessary.
Battery Aged
L
C
Replace the battery.
Battery failed charging
M
C
Battery failure or power board hardware failure. Replace
the battery. If the problem persists, contact your service
personnel.
Therapy module
Therapy Equip Selftest
Err
S
C
An error occurred during therapy module self test. Restart
the equipment or replace the therapy module low voltage
board.
Defib Malfunction
S
C
The defibrillation function fails or both the defibrillation
and pacing functions fail. Restart the equipment and test
defibrillation function. If the problem persists, contact
your service personnel.
Pacer Malfunction
S
C
The pacing function fails. Restart the equipment and test
pacer function. If the problem persists, contact your
service personnel.
Disarming Failed
H
C
There is a problem with the therapy module disarming
circuit. Replace the therapy module low voltage board
and high voltage board.
Monitoring
module
Monitor Module
Selftest Err
H
C
An error occurred during MPM module power-on self test.
Replace the MPM module.
Monitor Module Reset
Err
H
C
MPM module reset abnormally. In this case, the MPM
module restores to default configuration. You can ignore
this problem.
Monitor Module
Voltage Err
L
C
The voltage of MPM module is abnormal. Replace the
MPM module.
Recorder
Recorder Init Err
L
A
Restart the equipment.
Recordhead
Overheated
L
A
The recorder has been working for a prolonged time.
Clear the recording tasks and resume the recording till the
recorder’s print head cools down.
Recorder Overcurrent
L
A
Re-load the recorder paper.
Pacer
Pads Cable Off
H
C
Check that pads cable is properly connected.
Pads Off
H
C
Check that electrode pads are properly connected.
ECG Lead Off
H
C
Check that ECG leadwires are properly connected.
Pacer Stopped
Abnormally
H
C
Check that electrode pads well contact with patient’s skin.
Make sure electrode pads are properly applied, and then
start pacing again.
Others
Load Config Err
L
A
Check if the configuration is correct, or restore the factory
configuration.
Measurement
Alarm Message
L
I
Cause and solution
Содержание BeneHeart D2
Страница 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Страница 2: ......
Страница 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 204: ......
Страница 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Страница 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Страница 207: ......
Страница 208: ...P N 046 010599 00 9 0...