B-1
B
EMC
The equipment meets the requirements of IEC 60601-1-2.
NOTE
Use of accessories, transducers, and cables other than those specified may result in increased emission
and/or decreased electromagnetic immunity of the defibrillator/monitor.
The equipment or its components should not be used adjacent to or stacked with other devices. If adjacent
or stacked use is necessary, the equipment should be observed to verify normal operation in the
configuration in which it will be used.
The equipment needs special precautions regarding EMC and needs to be installed and put into service
according to the EMC information provided below.
Other devices may affect this monitor even though they meet the requirements of CISPR.
When the inputted signal is below the minimum amplitude provided in technical specifications, erroneous
measurements could result.
Portable and mobile RF communications equipment may affect this equipment.
Guidance and Declaration - Electromagnetic Emissions
The equipment is suitable for use in the electromagnetic environment specified below. The customer or the user of the
equipment should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The equipment uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
Meet the requirements of Class B
Guidance and Declaration - Electromagnetic Immunity
The equipment is suitable for use in the electromagnetic environment specified below. The customer or the user of the
equipment should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the
relative humidity should be at least 30%.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
Содержание BeneHeart D1
Страница 1: ...BeneHeart D1 Automated External Defibrillator Operator s Manual...
Страница 2: ......
Страница 8: ...VI FOR YOUR NOTES...
Страница 42: ...6 6 FOR YOUR NOTES...
Страница 50: ...7 8 FOR YOUR NOTES...
Страница 58: ...9 6 FOR YOUR NOTES...
Страница 86: ...B 4 FOR YOUR NOTES...
Страница 88: ...C 2 FOR YOUR NOTES...
Страница 92: ...D 4 FOR YOUR NOTES...
Страница 96: ...E 4 FOR YOUR NOTES...
Страница 100: ...F 4 FOR YOUR NOTES...
Страница 102: ...G 2 FOR YOUR NOTES...
Страница 104: ......
Страница 105: ......
Страница 106: ...P N 046 004673 00 3 0...