10-2
10.2
Battery Alarms
10.2.1
Low Battery Alarm
If the battery charge is low, a technical alarm “Low Battery” will be triggered. In this case, replace the battery with a fully
charged rechargeable battery or a new disposable battery on immediately.
If the battery is almost depleted, a prompt “Battery Depleted! Replace Battery Now.” pops up and alarm tones are
provided. In this case, replace the battery immediately. This prompt will not disappear until the battery is replaced. The
equipment automatically shuts down if no action is taken in 3 minutes.
NOTE
The Low Battery alarm means that the battery is beginning to weaken and should be replaced at the first
opportunity. At least 20 minutes of monitoring and 10 200J shocks can be performed when the Low
Battery alarm is activated. Replace the battery as soon as possible.
10.2.2
Battery Aged Alarm
If the battery runtime is significantly shorter than the specification, a low level technological alarm “Battery Aged” will be
presented. We recommend you to contact our company and replace it with a new one.
10.2.3
Battery Error Alarm
In the situation that the battery has a failure, a high level technological alarm “Battery Err” will be presented. In this case,
replace the battery or contact your service personnel.
Содержание BeneHeart D1
Страница 1: ...BeneHeart D1 Automated External Defibrillator Operator s Manual...
Страница 2: ......
Страница 8: ...VI FOR YOUR NOTES...
Страница 42: ...6 6 FOR YOUR NOTES...
Страница 50: ...7 8 FOR YOUR NOTES...
Страница 58: ...9 6 FOR YOUR NOTES...
Страница 86: ...B 4 FOR YOUR NOTES...
Страница 88: ...C 2 FOR YOUR NOTES...
Страница 92: ...D 4 FOR YOUR NOTES...
Страница 96: ...E 4 FOR YOUR NOTES...
Страница 100: ...F 4 FOR YOUR NOTES...
Страница 102: ...G 2 FOR YOUR NOTES...
Страница 104: ......
Страница 105: ......
Страница 106: ...P N 046 004673 00 3 0...