7-7
7.6.4
Automatic Arrhythmia Relearn
Arrhythmia relearning is initiated automatically whenever:
The ECG lead or lead label is changed
The ECG lead is re-connected
Patient category is changed
The paced status is changed,
Arrhythmia analysis is switched on
[
Stop Calibrating
] is selected after ECG calibration is completed.
NOTE
Arrhythmia relearning in the case of ventricular tachycardia may affect correct arrhythmia alarm.