C-2
Alarm
Information
Alarm
Level
Reason
Countermeasure
[No Battery]
High
1. Only powered with built-in
battery, battery is empty.
2. Built-in battery and external
DC power encounter failure.
Connect to the AC power
source to cancel alarm.
[VTBI Complete]
High
Infusion volume reaches the
preset VTBI.
1. Press
to cancel
alarm.
2. Or alarm is cancelled
when reach the KVO
infusion time.
3. Or press
to
confirm KVO infusion to
cancel alarm.
[KVO Finish]
High
KVO infusion has run for 30
minutes. Infusion is stopped
after the alarm is triggered.
Press
to cancel
alarm.
[Door opened]
High
The infusion pump door is
opened during infusion.
Infusion is stopped after the
alarm is triggered.
Press
to cancel
alarm and close the door
correctly.
[System Error]
High
Power supply voltage
abnormal (overhigh), AD value
of sensor abnormal,
communication abnormal and
other errors.
Alarm cannot be
cancelled. Please stop
operation and contact the
company.
[Empty]
High
If drop sensor is installed
correctly and the switch of
[Drop sensor] is on, no liquid
drop.
Note: If drop sensor is not
correctly inserted or the
surface of the liquid in the drip
chamber is abnormal, the
alarm might be triggered.
Infusion is stopped after the
alarm is triggered.
Press
to cancel
alarm, check and eliminate
the source of the alarm.
[No Infusion
Tube]
High
Infusion set is not loaded and
start infusion when the door is
closed.
Press
to cancel
alarm, correctly load
infusion set and close the
door.