7-3
please refer to the instructions of CIMS. Normal communication of the pump and
CIMS depends on whether the network connection is successful, operators are
unable to observe the operation status of the pump in real time when the
communication is interrupted. After the network connection settings of the pump and
CIMS are modified, operators shall reset the network connection as required in the
manual to ensure the communication of the pump and CIMS are restored.
NOTE
Wireless security transmission distance is no more than 50 meters.
2.4 GHz Wi-Fi frequency range, WEP and WPA/WPA2 security modes and
802.11b/g/n wireless standard are supported.
The settings of the wireless network must be conducted by technicians
approved by the company or maintenance staff designated by the
company.
It is recommended that not access the pump to public network. If the pump
is required to access to public network, firewall shall be added to prevent
cyber attacks.
7.5 Data Export
To export the data in the infusion pump, please refer to the following steps:
1.
Log on PC tools, and connect the PC to the infusion pump;
2.
When the infusion pump is in working communication with the PC, the PC
automatically reads all the data in the pump;
3.
Select [
History Record
] in PC tools, and export data.
7.6 Relay Setting
Multi-syringe pumps/infusion pumps have been inserted to BeneFusion DS5 Infusion
Supervision System and the pumps are on, follow the steps below to start relay
mode:
1. Select [
Main Menu
]→[
Select Mode
], then select [
Relay setting
] in each mode.
2. Set the slot No. of the “Forward channel”, press
to save. If the slot of
“Forward channel” pump is infusion pump, please input VTBI parameter value of
the pump.
3. When the upstream channel finishes infusion, the downstream channel with
will start running automatically. When the last relayed pump finishes infusion,
[
VTBI Complete
] alarm will be triggered.