1-3
Infusion pump with specified infusion sets has to be used with specified
infusion sets.
When operating the pump and under non-running status, please carefully
check whether there is leakage from the tube export of the infusion set,
and drop needle of the drip chamber. If any leakage is found, please
contact the manufacturer for maintenance and confirmation.
Do not touch the patient when connecting the peripheral equipment via the
input/output signal ports to prevent patient leakage current from exceeding
the requirements specified by the standard.
In the process of defibrillation, do not touch patient and other
non-defibrillation equipments to prevent electric shock damage, and
defibrillation will not affect the basic performance (such as infusion
accuracy, alarm and signal transmission) of the pump.
The equipment use a mains plug as isolation means to the mains power.
Do not locate the equipment in a place difficult to operate the mains plug.
1.1.2 Cautions
CAUTION
Use the accessories specified in this Operator’s Manual to guarantee the
patient’s safety.
When this infusion pump and its accessories exceed their service life, they
must be disposed of in accordance with local statutes or hospital
regulations. If you have any queries, please contact your distributor or the
manufacturer.
After loading the infusion set and before infusion, check for leakages. If
any are found, they should be rectified as soon as possible.
Infusion pump with conventional infusion sets: For SK series infusion
sets, it is recommended to replace the infusion set or adjust the fixing site
of the infusion set after the infusion has been running for 24 hours to
guarantee accuracy. For infusion sets of other brands, it is recommended
to test the service life of the infusion set to determine the time interval of
changing the fixing site of the infusion set; if the service life of the infusion
set is not tested, it is recommended to adjust the fixing site of the infusion
set every 4 hours after infusion begins to guarantee accuracy, and reload
the new tube without extrusion.
Infusion pump with specified infusion sets: It is recommended that the
time interval of changing infusion sets should be less than 72 hours.
Electromagnetic fields may affect equipment performance. This makes it