7-1
7
Other Functions
7.1 History Record
The infusion pump when in use will produce some key data stored in [
History
Record
], providing foundation for the treatment review and maintenance review at a
later period. The attribute of recording events includes action, time and description.
A record is created once an event occurs. The memory can store up to 2000 records.
Once the memory is full, the oldest records will be removed first. History record will
not loss when the infusion pump powers down.
1. Select [
Main Menu
]→[
History Record
].
2. Select [
History Record
]: Each page can demonstrate up to 3 records, and press
to turn the pages.
7.2 Power-down Save
To prevent the loss of patient data and alarm settings when the infusion pump
suddenly powers down, the infusion pump provides the function of the power-down
data and alarm settings storage. If the infusion pump powers down suddenly after it is
restarted, the last infusion parameters will display the alarm information and will
remain in consistency with those before the power-down, and will be reloaded.You
can refer to [
History Record
] to view such information as infusion parameters and
alarm.
7.3 Nurse Call
Select [
Main Menu
]→[
System Options
]→[
Nurse call
] , and set in the open menu:
Switch
On: Indicates switching on the nurse call function.
Off: Indicates switching off the nurse call function.
Signal type
Continuous: Indicates that the output nurse call signal type is the same as that of the
alarm existence time, i.e., from the occurrence of the alarm to the end of it.
Pulse: Indicates the output nurse call signal is a pulse signal with the type of 1 second.
When several alarms exist at the same time, only one pulse signal can be outputted.
If the current alarm is not removed and another alarm occurs, then one additional
pulse signal is outputted.
Trigger type
Normally Close: Select when the hospital call system is set as [