A-13
Unit of Rate (gtt/min)
Sampling rate: 20gtt/min
Sampling interval:
△
t =0.5min
Test period: t =120mins
Infusion rate: Q (gtt/min)
Sampling rate: 20gtt/min
Sampling interval:
△
t =0.5min
Observation windows: p
△
t = 2,
5, 11, 19, 31 mins
Maximum deviation over the
course of a full observation
window: Ep(Max) (%)
Minimum deviation over the
course of a full observation
window: Ep(Min) (%)
Average deviation: A (%)
Sampling rate: 20gtt/min
Sampling interval:
△
t =0.5min
Observation windows: p
△
t = 2,
5, 11, 19, 31mins
Maximum deviation over the
course of a full observation
window: Ep(Max) (%)
Minimum deviation over the
course of a full observation
window: Ep(Min) (%)
Average deviation: B(%)