1-4
necessary for other equipment used in the vicinity of the pump to meet
EMC standards. Mobile phones, X ray and MRI equipment are all potential
interference sources because of their high-intensity electromagnetic
radiation.
Before the equipment is connected to the power supply, check that the
voltage and frequency of the power supply match the specifications on the
label or in this Operator’s Manual.
Please install and carry the equipment correctly to protect the equipment
from damage from drops, impacts, violent shaking or other external
mechanical forces. The equipment should be observed to verify normal
operation after fall, otherwise it cannot be used.
Disposable accessories must be disposed of after use in accordance with
the relevant hospital regulations.
Avoid direct sunshine, high temperatures and dampness.
Check the built-in battery before use to make sure it has sufficient power.
Recharge the battery if necessary.
The infusion set with the luer taper is recommended for use, which can
effectively prevent patients from under current caused by the occurrence
of the cannula to slip out when under tension.
1.1.3 Notes
NOTE
Install the equipment in a position where it can be easily accessed for
inspection, operation and maintenance.
Keep this Operator’s Manual near to the equipment for future ease of
reference.
The software was developed in compliance with IEC62304. The possibility
of hazards arising from software errors is minimized.
This Operator’s Manual describes the most complete functional
configuration of the equipment. The product you are using may not have
some of the settings or functions described herein.
Do not insert devices that are not specified by the manufacturer into the
multifunction interfaces.
During infusion, the infusion pump can accurately control the rate,
infusion volume and infusion time, and monitor the operation in real-time,
to effectively prevent over currents, under currents and instances of
backflow.
The device is not in touch with the drugs or patients directly. Thus, there is
no need to process Biocompatibility test on it.