A-9
A.5.2 Infusion Pump with Silicone Specified Infusion Sets
Occlusion Pressure
(mmHg)
Rate
(ml/h)
Time of Occlusion Alarm
(hh:mm:ss)
75
0.1
01:07:08
1
00:05:02
25
00:00:30
975
0.1
18:45:50
1
01:19:12
25
00:02:50
Pressure Setting
Rate
(ml/h)
Bolus Volume (ml)
75
25
<0.1ml
975
25
<0.15ml
NOTE
Test conditions:
FLUKE IDA4 PLUS tester
Infusion set brand and model: M-IS-000
Test temperature: 20±2
℃
Extension tube length: 1 meter
Occlusion alarm pressure, alarm delays and bolus volume may vary
depending on test conditions, temperature and tube length.
The above data are only typical values under normal test conditions. The
actual data may vary as test conditions change. Please refer to the test
data for the product you have purchased. Under the same standard
occlusion value and rate, the higher the value of the tested pressure is, the
longer the alarm time will be delayed.