3-8
3. After standard specimen 1 is successfully calibrated, “Please install standard
specimen 2, and press
for calibration” is prompted in the screen.
4. After installing standard specimen 2 as shown in the figure correctly, press
to calibrate standard specimen 2.
5. If the calibration succeeds, the message "
Calibration Completed!
" is displayed.
6. After the calibration is successful, the current outer diameter is displayed close to
that of standard specimen 2.
3.5 Running Test
General functions:
1. When pressing any key, the syringe pump should be able to act or react
normally according to the Instructions for Use, and abnormal operation should
not cause a crash.
2. According to the Instructions for Use, operate the syringe pump to check
whether the sound and light alarms at all levels are normal, whether the volume
control is normal, and whether the alarm mute works.
3. After operating the pump according to the Instructions for Use and triggering an
alarm, check whether the historical records can record the operation and the
Standard specimen 1 (Outer diameter 13.5mm)
Standard specimen 2 (Outer diameter 32mm)
The positioning post is close to the
front of the syringe pump.
Содержание BeneFusion SP3
Страница 1: ...BeneFusion SP3 BeneFusion SP3 Ex Syringe Pump Service Manual ...
Страница 2: ......
Страница 12: ...1 6 FOR YOUR NOTES ...
Страница 24: ...2 12 FOR YOUR NOTES ...
Страница 62: ...5 16 FOR YOUR NOTES ...
Страница 64: ...6 2 6 2 Front Cover Assembly 6 2 1 Exploded View ...
Страница 66: ...6 4 6 3 Pump Module Assembly 6 3 1 Exploded View ...
Страница 78: ...7 8 FOR YOUR NOTES ...
Страница 79: ......
Страница 80: ...P N 046 013115 00 3 0 ...