4-6
connection
4.The motor is
faulty.
5.Poor
connection exists
in the link of long
potentiometer -
sensor adapter
board - control
board
6.Long
potentiometer is
faulty.
7.The control
board is faulty.
motor.
4.Replace the
motor.
5.Reconnect the
link.
6.Replace the
sensor adapter
board.
7.Replace the
control board.
SE306
/
Motor Not Started
Actually
Under normal
circumstances, it
may not appear.
There may be
software defects.
Feedback to
R&D for
analysis.
SE307
/
Motor Speed Out of
Range
Under normal
circumstances,
the speed
calculated by the
software will not
exceed the
range. There may
be software
defects.
Feedback to
R&D for
analysis.
SE402
/
Pressure Sensor
Failure
1.Poor
connection or
wire damages
exist in the link of
pressure sensor -
pressure
detection board -
slider adapter
board - sensor
adapter board -
control board.
2.The pressure
sensor is faulty.
3.The pressure
detection board is
1.Reconnect the
link or replace
the wires.
2.Replace the
pressure sensor.
3.Replace the
pressure
detection board.
4.Replace the
control board.
Содержание BeneFusion SP3
Страница 1: ...BeneFusion SP3 BeneFusion SP3 Ex Syringe Pump Service Manual ...
Страница 2: ......
Страница 12: ...1 6 FOR YOUR NOTES ...
Страница 24: ...2 12 FOR YOUR NOTES ...
Страница 62: ...5 16 FOR YOUR NOTES ...
Страница 64: ...6 2 6 2 Front Cover Assembly 6 2 1 Exploded View ...
Страница 66: ...6 4 6 3 Pump Module Assembly 6 3 1 Exploded View ...
Страница 78: ...7 8 FOR YOUR NOTES ...
Страница 79: ......
Страница 80: ...P N 046 013115 00 3 0 ...