3-3
accessories of the pressure sensor.
Running Test
At least once a year, or when you suspect an abnormality
in infusion flow, alarm pressure, etc.
Battery Powered Test
1.At least once a year.
2.After the battery of the syringe pump is repaired or when
the user suspects that the battery is the source of the fault.
Electrical
Safety
Tests
Enclosure
leakage current
test
1.After any repair or replacement of the power module.
2.After the equipment has dropped or has been seriously
collided.
Patient Leakage
Current Test
Cleaning and Disinfection
1.Once a month recommended.
2.Thorough cleaning of the syringe pump surface is
required before or after long storage.
3.2 Overall Appearance Inspection
1) The enclosure is clean and tidy, without scratches, and the assembly is tight, and
there is no abnormal noise when the equipment is shaken.
2) The keys works properly.
3) The labels are complete and the contents are correct.
4) The standard configuration is complete and the socket is installed tightly.
3.3 Power-on test
1) Press the power switch. The system performs self-test, and the [System
Self-Test] screen is displayed.
2) First, the speaker gives a “du” sound – the speaker self-test is successful.
3) Next, the buzzer emits two “di-di” sounds – the buzzer self-test is successful.
4) The alarm lamp lights up from red to yellow and then goes out - the alarm lamp
self-test is successful.
Note: If the power-on self-test fails, press
to enter the system, and then query
the history records and investigate the failure reason.
Содержание BeneFusion SP3
Страница 1: ...BeneFusion SP3 BeneFusion SP3 Ex Syringe Pump Service Manual ...
Страница 2: ......
Страница 12: ...1 6 FOR YOUR NOTES ...
Страница 24: ...2 12 FOR YOUR NOTES ...
Страница 62: ...5 16 FOR YOUR NOTES ...
Страница 64: ...6 2 6 2 Front Cover Assembly 6 2 1 Exploded View ...
Страница 66: ...6 4 6 3 Pump Module Assembly 6 3 1 Exploded View ...
Страница 78: ...7 8 FOR YOUR NOTES ...
Страница 79: ......
Страница 80: ...P N 046 013115 00 3 0 ...