5-11
3. As shown in the figure: Remove the hot melt adhesive on the clips (Note: if it is
difficult to remove, drip a drop of alcohol. Then you can remove it easily), release the
clips, pull out the FPC wire, tear the sponge pad, and then loosen the three M3X8
Phillips pan head screws that fix the PCBA. Finally remove PCBA from the pump
body.
4. As shown in the figure: Use a needle-nosed pliers to remove the collar on the
transition gear shaft and remove the transition gear from the pump body assembly.
5. As shown in the figure: Remove the five M3x12 Phillips pan head screws on the
pressure sensor holder with the Phillips screwdriver in the order shown in the figure.
Then remove the holder bracket and bearing.
Note: You should still follow the screw locking sequence as shown during assembly.
Tear the sponge pad
Pull out the
FPC wire
Transition Gear